Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
• Perform maintenance and repair work only on disinfected
operating tables! Disinfection is performed by the medical
facility.
• Perform work in the OR area only with appropriate
authorization.
• Immediately seek a physician in the event of a complaint of
possible infection. Inform the doctor you were working in a
hazardous area - even if this was several months prior.
Prevention against
infection
• Ask the doctor about possible risks and discuss how to avoid
them in your work as a medical service technician.
• Get a hepatitis B inoculation if advised by a physician. A booster
shot is required every 3 to 5 years. The incubation period for
hepatitis B is 1 to 6 months!
2.2
Liability
TRUMPF Medizin Systeme GmbH + Co. KG
IS LIABLE ONLY
for
reliable and proper functionality of the operating table
IF
• installation, modification, and repairs are performed by
Trumpf Medical service technicians or by personnel authorized,
trained, and certified by Trumpf Medical.
• the operating table is used properly in accordance with the
instruction manual.
2.3
Disposal
The products must be recycled in an environmentally friendly
manner. Disposal, including that of individual parts, must be
environmentally friendly, i.e., in accordance with the legal
regulations currently in force. For information on proper disposal of
old equipment, please contact either Technical Service at
Trumpf Medical, your local sales representative, or the appropriate
national agency Trumpf Medical will take back your old equipment
or products that are defective or no longer used. For detailed
information, contact Technical Service.
When decommissioning an operating table, the lithium ion batteries
have to be removed by a Trumpf Medical service technician or a
person trained and authorized by Trumpf Medical. Return removed
and unusable batteries in suitable packaging to Trumpf Medical
Technical Service. Important: the returns must be declared as
hazardous materials of class 9/UN3480! Trumpf Medical will
take responsibility for the environmentally proper disposal of the
battery.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......