Motor controller circuit board
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
57
7. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
8. Put on the pad.
9. Calibrate the tilt sensor (see the software description from the
CAN test center). Only necessary when installing a new motor
controller circuit board.
10. Configure the motor controller (only necessary when installing
a new motor controller circuit board). See the software
description from the CAN test center.
11. Update the software (see the software description from the
CAN test center).
TEST
Perform recheck according to IEC 60601-1; perform function test
and final check.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......