Cable W104
202
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
assembly (2 screws).
2. Note the original mounting position of the cable hold-down
clamp!
Do not jam cables! Stick the cable hold-down clamp onto the
two screws on the guide column, slide it carefully along the
oblong holes, press it down (cable clamp must rest correctly
on the battery) and attach it (tighten the 2 screws).
3. Connect cable W173 to W104 (coupling point on the
suspension plate).
4. Attach the ground wire connection of cables W173 and
W104 to the suspension plate (1 screw with 2 washers).
5. Restore the original cable routing. Note the shield support.
Refasten cable W104 to the suspension plate.
6. Insert the plug of cable W104 into the socket on the connector
board (ST2) (power supply socket) (see circuit diagram on
page 295).
7. Attach the ground wire connection of cable W104 to the
column base (1 screw with 2 washers).
8. Restore the original cable routing. Note the shield support.
Refasten cable W104 to the table base.
9. Set the cable deflector onto the two screws on the column
base and slide it flush against the power supply unit. Mount the
cable deflector to the power supply unit (2 screws on the top)
and tighten the two screws on the column base.
10. Attach frame piece 2 to frame piece 1 (2 screws from below).
11. Attach frame piece 2 to the tilt left assembly (2 screws with
washers).
12. Attach the metal panel with the column keypad (see
13. Mount the bellows on the frame (see chapter 9 on page 31).
14. Slide the column cover up and close it (see chapter 9 on
page 31).
15. Open the column cover at the bottom to establish the power
supply (see chapter 8 on page 30).
16. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
17. Put on the pad.
TEST
Perform recheck according to IEC 60601-1; perform function test
and final check.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......