Column controller circuit board
136
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
3. Connect the column controller to the plug contact (W164) on
the Trendelenburg assembly.
4. Attach the column controller to the tilt right assembly (1 screw
with washer).
5. Mount the bellows on the frame (see chapter 9 on page 31).
6. Attach the metal panel with the column keypad (see
7. Slide the column cover up and close it (see chapter 9 on
page 31).
8. Open the column cover at the bottom to establish the power
supply (see chapter 8 on page 30).
9. Establish the internal power supply, close the column cover
and turn on the operating table (see chapter 10 on page 33).
10. Put on the pad.
11. Update the software (see the software description from the
CAN test center).
TEST
Perform recheck according to IEC 60601-1; perform function test
and final check.