Information on the hydraulics system
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
159
opened, no dirt gets into the valves. Keep openings on the hydraulic
system open only as long as it absolutely necessary, and then close
them tight. Use only lint-free materials to clean the hydraulic system.
The maximum pressure in the system is 160 bar. The head and foot-
end floor locks are extended with different pressure. The foot-end
floor locks compensate for unevenness in the floor. Note the head
and foot arrangement of the locking cylinder in the table base. The
head-end floor locks (lift 25) have two wrench flats, and the foot-
end floor lock (lift 35) has one wrench flat.
The hydraulic hoses have to be routed in the operating table base
with a bending radius of greater than 35 mm. Any less of a bending
radius and the hose could kink and damage the hydraulic system.
When working on the hydraulics unit, make sure no air seeps into
the system. Therefore, do not activate the hydraulics on the
operating table if the operating table has been placed on its side
with the column keypad pointing down.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......