Open and close the lower column cover
30
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
8
Open and close the lower column cover
Open
1. Prepare the operating table (see chapter 7 on page 29).
2. Turn off the operating table.
3. Disconnect the lowest column cover [1] from the pot [2] (2
screws [3]).
4. Carefully slide the column cover upward, tighten the retaining
screw [4] on the Trendelenburg assembly [5], and place the
column cover [6] on the retaining screw.
Close
1. Grasp the column cover, remove the holding screw from the
Trendelenburg assembly and guide the cover carefully
downward.
2. Install the lowest column cover on the pot (2 screws).
3. Put on the pad.
4. Switch on the operating table at the column keypad.
5. Connect the operating table to the external power supply.
CAUTION
Risk of injury due to capacitor voltage.
Switch off repeatedly if needed. No LEDs may be illuminated
on the column keypad!
TEST
Perform a function test.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......