Cable W172
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
215
21.7
Cable W172
Disassembly
1. Prepare the operating table (see chapter 7 on page 29).
2. Switch off the operating table and disconnect the internal
power supply (see chapter 10 on page 33).
3. Remove the cable deflector [1] from the power supply unit [2].
To do this, unscrew the 2 screws [4] at the top of the power
supply unit [2] and loosen the two screws [3] on the column
base. Guide the cable deflector out of the screws [3] on the
column base and remove it.
4. Remove the cover [5] over the coupling point on the side of
table base (2 screws).
5. Unplug the cable W172 [6] from the coupling point [9] in the
table base.
6. Pull both plugs at the other end of cable W172 [6] from the
power board (PB1) [7] on the power supply unit.
7. Note the cable routing and attachment.
Remove cable attachment along cable W172 [6] from
battery 2 [8] to the connector board [7] on the power supply
unit and from battery 2 [8] to the coupling point [9] on the
table base. Remove cable W172 [6].
Assembly
1. Restore the original cable routing.
Route cable W172 in the column pot.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......