Disconnecting and reconnecting the internal power supply on the operating table
Service Manual TruSystem 7000 Operating Table – 1 764 986 – 10/2016
33
10
Disconnecting and reconnecting the internal power supply on the
operating table
Disconnect the
power supply
1. Prepare the operating table (see chapter 7 on page 29).
2. Turn off the operating table.
3. Open the lower column cover (see chapter 8 on page 30).
4. Pull the power supply plug from battery 1 [1] and battery
2 [2].
Reconnect the power
supply
1. Connect the power supply plugs to battery 1 and battery 2
(see the circuit diagram on page 295). The plug must lock in.
2. Close the column cover (see chapter 8 on page 30).
3. Put on the pad.
4. Switch on the operating table at the column keypad.
5. Connect the operating table to the external power supply.
CAUTION
Risk of injury due to capacitor voltage.
Switch off repeatedly if needed. No LEDs may be illuminated
on the column keypad!
TEST
Perform a function test.
Summary of Contents for TruSystem 7000
Page 1: ...TruSystem 7000 Operating Table Service Manual...
Page 4: ...Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 10: ...Contents 10 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 297: ...Circuit Diagram Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016 297...
Page 298: ...Circuit Diagram 298 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 300: ...Hydraulics Diagram 300 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 302: ...Hydraulic Jack Unit 302 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 304: ...Service Information 304 Service Manual TruSystem 7000 Operating Table 1 764 986 10 2016...
Page 305: ......