Remote Control
R&S
®
NRT2
99
User Manual 1178.5550.02 ─ 04
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
0
0
NRT POWer
0
0
0
0
0
0
NRT SERRRor
0
0
0
Warning
0
+
Figure 11-3: Questionable status register
Querying the register:
●
STATus:QUEStionable:CONDition?
●
Table 11-2: Used questionable status bits and their meaning
Bit no.
Short description
Bit is set if
3
NRT POWer
●
Sensor is overloaded; the forward or reverse power
exceeds the maximum values specified for the sensor.
●
The measured values for power or matching exceed the
limit values.
●
The central supply voltage for power sensor and R&S
NRT2 exceeds or falls below the allowed limits.
10
NRT SERRRor
Static error of the R&S
NRT
‑
Zxx power sensor occurred.
14
Warning
11.2.6
Standard Event Status and Enable Register (ESR, ESE)
The
ESR
is already defined in the IEEE 488.2 standard. It is comparable to the
EVENt
register of a SCPI register. The standard event status register can be read out by
.
The
ESE
forms the associated
ENABle
register. It can be set and read by
Status Reporting System