![Rohde & Schwarz 1430.0105K02 User Manual Download Page 145](http://html1.mh-extra.com/html/rohde-and-schwarz/1430-0105k02/1430-0105k02_user-manual_1477792145.webp)
Remote Control
R&S
®
NRT2
145
User Manual 1178.5550.02 ─ 04
OUTPut:RECorder<output>:LIMit:LOWer[:POWer]
<lower_limit>
Specifies the lower power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power (unit
DBM
,
W
or
DBUV
) as measured
value.
Units:
DBM | W | DBUV
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
1e-15 to 1e12
*RST:
1e-6
Default unit: W
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer:RATio[:VALue]
<value>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power ratio (unit
DB
,
DPCT
or
O
) as the mea-
sured value.
Unit:
DB | DPCT | O
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
1e-18 to 1e18
*RST:
10.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:UPPer[:POWer]
<value>
Specifies the upper power limit of the characteristic for one of the two analog outputs if
the associated calculate block returns a power (unit
DBM
,
W
or
DBUV
) as the measured
value.
Unit:
DBM | W | DBUV
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
1e-15 to 1e12
*RST:
1.0
Default unit: W
Manual operation:
See
OUTPut