Remote Control
R&S
®
NRT2
160
User Manual 1178.5550.02 ─ 04
OUTBound
A high level is output if the measured power is out of the range
specified by the scale limits of the lefthand bargraph.
HIGH
A high level is output if the measured power exceeds the upper
scale limit of the lefthand bargraph.
*RST:
HIGH
Example:
SENS1:POW:RANG:LIM:DET INB
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
<state>
This command can be used to define the AUX TTL connector as a monitoring output
for the power indication (ON state). In OFF state, the connector may be defined either
as a monitoring output for the matching or as a trigger input.
If the AUX TTL connector is assigned more than one function, the error message –221
"Settings conflict“ is generated and the first setting retained.
Parameters:
<state>
Range:
"ON", "OFF"
*RST:
OFF
Example:
TRIG:SOUR INT
SENS1:POW:REFL:RANG:LIM OFF
SENS1:POW:RANG:LIM ON
Manual operation:
See
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
<lower>
This command defines the lower scale limit for the lefthand bargraph (power indica-
tion).
Since the entry is made without unit, the following should be observed:
●
The selected unit is decisive for the absolute value of the scale limit.
●
The entered numeric value remains the same when another unit is selected (W,
dBm, %, dB) so that a change of the unit also causes a change of the scale limit.
●
For the entered scale limit to become effective, the automatic scaling of the bar-
graph has to be switched off.
Parameters:
<lower>
Range:
-1999.0 to 1999.0
*RST:
0.0
Example:
A lower scale limit of 150 mW is set with the following com-
mands:
SENS1:POW:RANG:AUTO OFF
SENS1:POW:RANG:LOW 0.15
UNIT1:POW W
SENSe