![Rohde & Schwarz 1430.0105K02 User Manual Download Page 170](http://html1.mh-extra.com/html/rohde-and-schwarz/1430-0105k02/1430-0105k02_user-manual_1477792170.webp)
Remote Control
R&S
®
NRT2
170
User Manual 1178.5550.02 ─ 04
SYSTem:COMMunicate:INET[:SELF]:DNS:SUFFix
<Domain>
Stores a DNS suffix which can be used in name resolution queries.
Parameters:
<suffix>
Manual operation:
See
SYSTem:COMMunicate:NETWork[:COMMon]:DOMain
<domain>
Sets the domain of the network.
Parameters:
<domain>
Example:
:SYSTem:COMMunicate:NETWork:COMMon:DOMain
ABC.DE
Sets the domain of the network to
ABC.DE
.
Manual operation:
See
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
<hostname>
Sets the individual host name of the power sensor.
In a LAN that uses a DNS server (Domain Name System server), each instrument con-
nected in the LAN can be accessed via an unambiguous host name, instead of the IP
address. The DNS server translates the host name to the IP address. This is especially
useful when a DHCP server is used, as a new IP address may be assigned each time
the instrument is restarted.
The sensor performs the change of the hostname immediately after the command is
sent. For this purpose the sensor restarts its connection to the network. During this
time that may take several seconds the sensor cannot be addressed. After the restart
the sensor can be addressed only through the newly set hostname.
Note:
It is recommended that you do not change the default host name in order to
avoid problems with the network connection. However, if you change the host name be
sure to use a unique name.
Parameters:
<hostname>
Example:
SYSTem:COMMunicate:NETWork:COMMon:HOST
'powersensor-2nd-floor'
Sets the hostname to
powersensor-2nd-floor
.
Manual operation:
See
SYSTem:COMMunicate:NETWork[:COMMon]:WORKgroup
<Workgroup>
Sets an individual workgroup name for the instrument.
SYSTem