Remote Control
R&S
®
NRT2
124
User Manual 1178.5550.02 ─ 04
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.5
Default unit: -
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
CALCulate<Measurement>:RELative<DirectionalChannel>:RATio[:VALue]
<value>
Relative value for power ratio.
Suffix:
<Measurement>
.
1...100
<DirectionalChannel> 1..2
Parameters:
<value>
Range:
1e-18 to 1e18
*RST:
1.0
Default unit: -
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
CALCulate<Measurement>:RELative<DirectionalChannel>[:STATe]
<state>
DirectionalChannel
means Forward and Reverse.
Suffix:
<Measurement>
.
1...100
<DirectionalChannel> 1..2
Parameters:
<state>
OFF | ON | SET
*RST:
OFF
Manual operation:
See
"Forward Relative State, Reflection Relative State"
CALCulate<Measurement>:RESolution
<resolution>
Selects the result resolution. This affects both the display resolution as well as resolu-
tion settings for the sensor(s).
Suffix:
<Measurement>
.
1...100
Parameters:
<resolution>
I | OI | OOI | OOOI
*RST:
OOI
CALCulate