Remote Control
R&S
®
NRT2
152
User Manual 1178.5550.02 ─ 04
Parameters:
<integration_time>
Range:
8.3e-9 to 30.0
*RST:
0.005
Default unit: s
[SENSe<Sensor>:]FREQuency[:CW]
<frequency>
This command informs the NRT of the carrier frequency of the specified sensor.
The entered frequency value is required to correct the frequency-dependent response
of the sensor. For this purpose NRT sensors contain a ROM with all sensor-specific
data. For NAP sensors frequency-dependent calibration factors can be entered via the
CALibration system.
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
1.0e9
Default unit: Hz
Example:
SENS1:FREQ 10 MHz
Manual operation:
See
See
[SENSe<Sensor>:]FREQuency:FIXed
<frequency>
This command informs the NRT of the carrier frequency of the specified sensor.
The entered frequency value is required to correct the frequency-dependent response
of the sensor. For this purpose NRT sensors contain a ROM with all sensor-specific
data. For NAP sensors frequency-dependent calibration factors can be entered via the
CALibration system.
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
1.0e9
Default unit: Hz
Example:
SENS1:FREQ 10 MHz
SENSe<Sensor>:BURSt:MODE
<mode>
This command instructs the NRT how to measure the average burst power.
Parameters:
<mode>
AUTO | USER
AUTO
The sensor automatically recognizes the duty cycle of the burst
series and calculates the average burst power from this duty
cycle and the average power. An appropriate video bandwidth
has to be set . For NAP-sensors, AUTO cannot be selected.
SENSe