Remote Control
R&S
®
NRT2
137
User Manual 1178.5550.02 ─ 04
Example:
MMEM:CAT:LENG? 'E:\'
Returns the number of files at the highest directory level of the
memory stick.
1
Response: There is 1 file at the highest directory level of the
memory stick.
Usage:
Query only
MMEMory:CDIRectory
[<directory_name>]
MMEMory:CDIRectory?
[<directory_name>]
Changes the default directory. This directory is used for all
MMEM
commands if no path
is specified. It is also possible to change to a higher directory using two dots '..' .
Parameters:
<directory_name>
*RST:
D:\
Example:
MMEM:CDIR 'test'
Changes from the current directory level to the
test
subdirec-
tory.
MMEMory:COPY
<file_source>[, <file_destination>]
Copies the specified file or directory (first parameter) to the specified destination (sec-
ond parameter).
Setting parameters:
<file_source>
<file_destination>
If the file destination not specified, <file_source> is copied to the
default directory on the default drive.
Files with the same name that already exist in the destination
directory are overwritten without an error message.
Example:
MMEM:COPY 'D:\user\test1.savrcl','E:'
Copies the
test1.savrcl
file in the
user
directory on the
D:
drive to the memory stick without changing the file name.
Usage:
Setting only
MMEMory:DATA
<file_name>, <data>[, <append>]
MMEMory:DATA?
<file_name>
Writes the <data> to the file identified by <file_name>. The IEC/IEEE bus terminator
should be set to EOI in order to ensure correct data transfer.
The associated query transfers the specified file from the R&S NRT2 to the control
computer. Make sure that the intermediate memory on the control computer is large
enough to receive the file. In this case, the setting for the IEC/IEEE bus terminator is
irrelevant.
MMEMory