Remote Control
R&S
®
NRT2
173
User Manual 1178.5550.02 ─ 04
SYSTem:INFO[:INFO]?
[<argument>]
Returns information about the system.
If queried without parameters, the command returns all available information in the
form of a list of strings separated by commas.
If you want to query specific information, add the query parameter:
SYST:INFO? "<string>"
Query parameters:
<argument>
'Manufacturer', 'Type', 'Stock Number', 'Serial', 'SW Build', 'MAC
Address', 'Hostname', 'IP Address', 'Domain', 'Subnetmask', 'Ga-
teway', 'Mode', 'Status', 'Sensor Name', 'Technology', 'Function',
'MinPower', 'MaxPower', 'MinFreq', 'MaxFreq', 'Impedance',
'Coupling', 'Uptime', 'Cal. Misc.', 'Cal. Abs.', 'Cal. Refl.', 'Cal.
Temp.', ' Cal. Lin.', 'Cal. S-Para.', 'Cal. S-Para. (User)', 'SPD
Mnemonic', 'Cal. Due Date', 'Certificate No', 'Limit', 'TestLimit',
'TestLimit pd'
Usage:
Query only
Manual operation:
See
SYSTem:IDN:MODE
<mode>
Selects if the automatically created instrument identification string is used or a user-
defined option string can be created and used.
Parameters:
<mode>
AUTO | USER
AUTO
The automatically instrument identification string is used.
USER
The user-defined instrument identification string is used. The
user-defined instrument identification string is defined with com-
mand
*RST:
AUTO
Manual operation:
See
SYSTem:IDN:ANSWer
<string>
Defines the return value for the
*IDN
query. The maximum string length is 128 charac-
ters.
Parameters:
<string>
The string entered is returned with query
*IDN?
.