Remote Control
R&S
®
NRT2
146
User Manual 1178.5550.02 ─ 04
OUTPut:RECorder<output>:LIMit:LOWer:CCDF
<value>
Specifies the lower limit of the characteristic for one of the two analog outputs. Below
this limit, the output is set to 0 V.
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
0.0
Default unit: pct
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RCOefficient
<value>
Specifies the lower limit of the characteristic for one of the two analog outputs. Below
this limit, the output is set to 0 V.
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
0.0
Default unit: -
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RFRatio
<value>
Specifies the lower limit of the characteristic for one of the two analog outputs. Below
this limit, the output is set to 0 V.
Suffix:
<output>
.
1...2
Parameters:
<value>
Range:
-1e18 to 1e18
*RST:
0.0
Default unit: pct
Manual operation:
See
OUTPut:RECorder<output>:LIMit:LOWer:RATio:RLOSs
<value>
Specifies the lower limit of the characteristic for one of the two analog outputs. Below
this limit, the output is set to 0 V.
Suffix:
<output>
.
1...2
OUTPut