![Rohde & Schwarz 1430.0105K02 User Manual Download Page 191](http://html1.mh-extra.com/html/rohde-and-schwarz/1430-0105k02/1430-0105k02_user-manual_1477792191.webp)
Troubleshooting
R&S
®
NRT2
191
User Manual 1178.5550.02 ─ 04
In remote control, notifications and errors are associated with a unique number. Posi-
tive numbers are instrument-dependent. Negative numbers are reserved by the SCPI
standard.
The notification types are grouped in number ranges. In the following description, both
the number used in remote control and the description is given to help identify the
problem.
12.2.1.1
System Notifications
Number range: 1000 to 1999. Described notifications:
.......................................................................................191
................................................................................................ 191
1007 - target descriptor error
...................................................................................... 191
..............................................................................................191
1004 - firmware update error
Firmware update failed.
Possible reasons:
●
You have used an
*.rsu
file that is not designated for the R&S NRT2. The name
of a suitable
*.rsu
file starts with "NRT2".
●
The firmware update was interrupted or otherwise faulty.
Solution: Perform the firmware update again. See
1005 - settings conflict
Settings conflict of the R&S NRT2 occurred.
Reason: Contradictory settings are allowed so that you are not hampered in your work-
flow.
Solution: Solve the settings conflict before starting a measurement.
1007 - target descriptor error
Servicing required. You cannot resolve the problem yourself.
Solution: Contact customer support. See
Chapter 12.5, "Contacting Customer Sup-
1008 - temperature alert
R&S NRT2 is overheated. Overheating can damage the R&S NRT2.
Possible reasons:
●
Insufficient airflow. Follow the instructions in
Chapter 3.1.4, "Setting Up the Prod-
●
The environmental temperature exceeds the suitable temperature range given in
the data sheet under environmental conditions.
●
The fan does not work properly. See
Chapter 12.5, "Contacting Customer Support"
Notifications