Getting Started
R&S
®
NRT2
19
User Manual 1178.5550.02 ─ 04
3.1.7
Connecting to LAN
See
Chapter 9.1.1, "Network Settings"
3.1.8
Connecting Power Sensors
The R&S NRT2 supports the R&S
NRT
‑
Zxx directional power sensors listed in the data
sheet. You have two choices for connecting the power sensors, but only one measure-
ment at a time is possible.
3.1.8.1
NRT Sensor Connector
See
.
Communication between the R&S
NRT
‑
Zxx directional power sensor and a base unit is
only possible with a baud rate setting of 38400
Bd. This setting is the factory default
that must be restored if the setting was changed. If the R&S
NRT
‑
Zxx directional power
sensor is not recognized by the base unit, check that the baud rate setting of the
R&S
NRT
‑
Zxx directional power sensor is 38400
Bd.
See the manual of the R&S
NRT
‑
Zxx directional power sensor for details.
The arrow on the power sensor casing shows the forward power flow.
1
2
3
4
5
6
Figure 3-1: Connecting to source and load
1 = source
2 = port
1 (RF connector)
3 = R&S
NRT
‑
Zxx directional power sensor
4 = port
2 (RF connector)
5 = load
6 = host interface connector
To connect the R&S
NRT
‑
Zxx directional power sensor
Connect the R&S
NRT
‑
Zxx directional power sensor between source and load of your
test setup as follows.
Preparing for Use