INPut<Sensor>:PORT:SOURce:AUTO
<auto>
This command switches the automatic assignment of the forward direction on or off.
Parameters:
<auto>
ON
With automatic assignment of the forward direction, the direction
in which the greater power flows is taken as the forward direc-
tion.
OFF
If the automatic assignment is switched off, e.g. to obtain a clear
assignment in the case that forward and reverse power are
almost equal, the orientation of the sensor is to be defined with
the
INPut<Sensor>:PORT:SOURce
command.
*RST:
0
Example:
INP2:PORT:SOUR:AUTO OFF