![Rohde & Schwarz 1430.0105K02 User Manual Download Page 184](http://html1.mh-extra.com/html/rohde-and-schwarz/1430-0105k02/1430-0105k02_user-manual_1477792184.webp)
Remote Control
R&S
®
NRT2
184
User Manual 1178.5550.02 ─ 04
Usage:
Query only
11.16
TEST
................................................................................................184
TEST:SENSor<Sensor>?
This command returns the status of the active sensor.
Example:
TEST:SENS ?
The response is sensor-dependent and could read as follows:
"NRT-Z44 V1.40","TEMPERATURE ERR", "CAL VALUES CHECKSUM ERR"
The response always contains an identification string for the
sensor with the type name and the FW version number. In case
of an error, error messages in plain text are appended. The con-
tent of the response string is identical with the status query in
manual control.
Usage:
Query only
Manual operation:
See
TEST:ALL?
The command tests all memories of the instrument. This command is a query and
therefore has no *RST value.
Parameters:
<sensor_command> *RST:
-
Example:
TEST?
Usage:
Query only
TEST:DIRect
<sensor_command>
This command is used for transferring direct setting commands to the active sensor.
The setting commands are contained in the sensor manual. With a ? appended, the
command returns the response of the sensor to the transferred command.
This command has no *RST value.
Parameters:
<sensor_command> *RST:
-
TEST