Remote Control
R&S
®
NRT2
175
User Manual 1178.5550.02 ─ 04
Manual operation:
See
SYSTem:OPT:AUTO
<state>
Activates/deactivates the return of a user-defined string for the
*OPT
query.
Parameters:
<status>
ON | OFF
*RST:
1
Manual operation:
See
SYSTem:SPEed
<mode>
The data processing speed of the R&S NRT2 can be increased when
FAST
is
selected. The display is switched off and the measured values are no longer displayed
since the continuous update of the screen content requires computation time.
Parameters:
<mode>
NORMal | FAST
*RST:
NORMal
SYSTem:PRESet
Triggers a sensor reset.
Usage:
Event
SYSTem:REBoot
Reboots the power sensor.
Usage:
Event
SYSTem:ERRor:ALL?
Queries all unread entries in the error/event queue and removes them from the queue.
The response is a comma-separated list of error numbers and a short description of
the error in the first in first out order.
Positive error numbers are instrument-dependent. Negative error numbers are
reserved by the SCPI standard.
Usage:
Query only
SYSTem:ERRor:CODE:ALL?
Queries all unread entries in the error/event queue and removes them from the queue.
Only the error numbers are returned and not the entire error text.
SYSTem