Remote Control
R&S
®
NRT2
138
User Manual 1178.5550.02 ─ 04
This command can be used to read or transfer stored instrument settings or waveforms
directly from/to the instrument.
Parameters:
<data>
<block>
The binary data block has the following structure:
#234<block_data>
#
Always comes first in the binary block.
<number>
Indicates how many digits the subsequent length entry has (2 in
example).
<number>
Indicates the number of subsequent bytes (34 in example).
<block data>
Block data for the specified length.
*RST:
-
Setting parameters:
<append>
APPend
Parameters for setting and query:
<file_name>
Example:
"MMEM:DATA 'TEST1.WV',#3767<block_data>"
Writes the block data to the
test1.wv
file.
MMEM:DATA? 'TEST1.WV'
Sends the data of the
Test1.wv
file from the R&S NRT2 to the
control computer in the form of a binary block.
MMEMory:DCATalog?
[<path_name>]
Returns the subdirectories of the specified directory in a list. The subdirectory names
are separated by commas. If no directory is specified, the default directory is read out.
Query parameters:
<path_name>
Return values:
<file_entry>
Response string: subdirectory name
Example:
MMEM:DCAT?
Returns the subdirectories of the default directory.
'test', 'wave', 'digital'
Response: The
test
,
wave
and
digital
subdirectories exist
in the default directory.
Usage:
Query only
MMEMory