Remote Control
R&S
®
NRT2
154
User Manual 1178.5550.02 ─ 04
[SENSe<Sensor>:]DM:STATe
<state>
This command switches the correction of the measurement value for modulated sig-
nals on or off. Selection of a communication standard, see
on page 153, is enabled only if the correction is switched on.
Parameters:
<state>
Range:
"ON", "OFF"
*RST:
0
Example:
SENS1:DM:STAT ON
Manual operation:
See
[SENSe<Sensor>:]DM:WCDMa:CRATe
<value>
This command defines the chip rate for the communication standard WCDMA. The
allowed range is sensor-dependent; the value is always entered without the unit (s
-1
).
Parameters:
<value>
Range:
0.0 to 8.2e6
*RST:
1.0e6
Default unit: Hz
Example:
SENS1:DM:WCDM:CRAT 4.096E6
[SENSe<Sensor>:]DATA?
[<function>]
This command provides access to the results of the current measurement(s). It is pos-
sible to request the result of a certain measurement function or to query the results of
all measurement functions switched on. The measurement functions are configured
with the
SENS<n>:FUNC...
commands.
The responses are output in the unit selected for the corresponding function. For some
functions, the unit can be selected.
By defining the measurement function, the result of a special measurement function is
requested. Without this specification the results of all active functions are queried. The
responses are given in the sequence of the measurement functions listed under
[SENS<n>:]FUNC:ON
.
The command is a query and therefore has no *RST value.
Query parameters:
<function>
Example:
SENS2:DATA? "POW:FORW:AVER"
Returns the measured result for the average forward power.
Usage:
Query only