System Settings
R&S
®
NRT2
70
User Manual 1178.5550.02 ─ 04
Figure 9-2: Example
Mode ← I/O 1, I/O 2 tabs
Sets the functionality of the Out
1
/
Trig
Out and Trig
In
/
Out
2 connectors.
"Off"
Disables the connector.
"Forw Analog Out", "Refl Analog Out"
Provides an analog voltage that is proportional to the displayed value.
"Forw Limit Violation"
Available for:
●
Out
1
/
Trig
Out BNC connector ("I/O
1" tab)
Sets the fail voltage that is output if a value of the forward measure-
ment causes a limit violation.
"Refl Limit Violation"
Available for the Out
1
/
Trig
Out BNC connector ("I/O
1" tab).
Sets the fail voltage that is output if a value of the reflection measure-
ment causes a limit violation.
Remote command:
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit[:STATe]
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LIMit[:STATe]
OUTPut:MODE<output>
0 V Equivalent ← I/O 1, I/O 2 tabs
Available if
is set to:
●
"Forw Analog Out"
●
"Refl Analog Out"
Enter the measurement value that corresponds to 0
V output voltage.
Connections