Remote Control
R&S
®
NRT2
153
User Manual 1178.5550.02 ─ 04
USER
The duty cycle is defined by the burst width and burst period,
see
on page 153 and
on page 153. The NRT calcu-
lates the average burst power from this duty cycle and the aver-
age power.
*RST:
AUTO
Example:
SENS1:BURS:MODE AUTO
SENSe<Sensor>:BURSt:PERiod
<value>
This command defines the burst period. It is only valid in the "USER" mode, see
Parameters:
<value>
Range:
0.0 to 1.0
*RST:
0.1
Default unit: s
Example:
SENS1:BURS:PER 1ms
SENSe<Sensor>:BURSt:WIDTh
<width>
This command defines the width of a burst. It is only valid in the "USER" mode, see
Parameters:
<width>
Range:
0.0 to 1.0
*RST:
0.01
Default unit: s
Example:
SENS1:BURS:WIDT 10ms
[SENSe<Sensor>:]DM:STANdard
<standard>
This command selects the communication standard. Which settings are available
depends on the sensor used. The settings are valid only if the modulation correction is
switched on via the
INPut<n>:DM:STATe ON
command.
Parameters:
<standard>
IS95 | WCDMa | DVBT | DAB
*RST:
IS95
Example:
SENS1:DM:STAN IS95
Manual operation:
See
SENSe