Remote Control
R&S
®
NRT2
161
User Manual 1178.5550.02 ─ 04
[SENSe<Sensor>:]POWer[:POWer]:RANGe[:UPPer]
<upper>
This command defines the upper scale limit for the lefthand bargraph (power indica-
tion). The entry is made without unit. For details, see
SENSe<Sensor>:POWer[:POWer]:RANGe:LOWer
.
Parameters:
<upper>
Range:
-1999.0 to 1999.0
*RST:
1.0
Example:
An upper scale limit of 35.7 dBm is set with the following com-
mands:
SENS1:POW:RANG:AUTO OFF
SENS1:POW:RANG 35.7
UNIT1:POW DBM
[SENSe<Sensor>:]POWer:REFerence
<ref>
Enters the reference values (in W or dBm) for the relative power indication (in %, dB,
or dBm). Units: Watt or dBm.
Parameters:
<ref>
Range:
0.0 to 100.0e6
*RST:
1.0
Default unit: W
Manual operation:
See
"Forward Reference Value, Reflection Reference Value"
[SENSe<Sensor>:]RRESolution
<rres>
This command changes the measurement resolution. This has an effect on the accu-
racy of the measurement, its duration, and on the number of digits indicated in the dis-
play.
Parameters:
<rres>
LOW | HIGH
*RST:
LOW
Example:
SENS1:RRES LOW
Manual operation:
See
[SENSe<Sensor>:]SWR:LIMit
<limit>
This command sets the limit value for the SWR at which an alarm is triggered. For an
alarm signal to be output, the forward power must also exceed a preset threshold
THReshold.
Parameters:
<limit>
Range:
1.0 to 100.0
*RST:
3.0
SENSe