Remote Control
R&S
®
NRT2
131
User Manual 1178.5550.02 ─ 04
CALCulate<Measurement>:LIMit:TYPE
<type>
The command allows selection of the value to be output.
Suffix:
<Measurement>
.
1...100
xxx
Parameters:
<type>
MAX | MIN | DIFFerence
MAXimum
Maximum value
MINimum
Minimum value
DIFFerence
Difference between maximum and minimum value
*RST:
MAX
Example:
CALC1:LIM:TYPE MAX
11.5
CALibration
........................................................................................... 131
CALibration<Sensor>:ZERO
This command causes zeroing to be carried out for the sensor connected to port <sen-
sor>. No RF power shall be applied to the sensor during zeroing.
Zeroing must be completed before a new measurement can be started. The command
*WAI
allows to recognize the end of a zeroing procedure.
Example:
CAL2:ZERO
Usage:
Event
11.6
DIAGnostic
...............................................................................................131
DIAGnostic:INFO:OTIMe?
This command returns the count of the built-in elapsed-time meter. The count is always
output in hours [h] and cannot be changed from outside.
The command is in the form of a query and therefore has no *RST value.
Example:
DIAG:INFO:OTIM?
DIAGnostic