![Rohde & Schwarz 1430.0105K02 User Manual Download Page 36](http://html1.mh-extra.com/html/rohde-and-schwarz/1430-0105k02/1430-0105k02_user-manual_1477792036.webp)
Operating Concepts
R&S
®
NRT2
36
User Manual 1178.5550.02 ─ 04
4.3.1
Switching to Remote Control (REMOTE)
Prerequisites
●
A link is established between the controller and the R&S NRT2.
●
The R&S NRT2 is configured correctly.
After power-up, the R&S NRT2 is always in manual control mode, "LOCAL". When the
R&S NRT2 receives a SCPI command, it switches to remote control irrespective of the
selected interface.
4.3.2
Returning to Manual Operation (LOCAL)
If the R&S NRT2 is in remote control, you can display settings using the front-panel
keys and the touchscreen, but you cannot change settings. To do that, you have to
return to manual operation.
The R&S NRT2 remains in remote control until you perform one of the following
actions. Make sure that the R&S NRT2 is free for you to use.
► Press the [Esc/Local] key. See
If the manual operation was disabled by the
&LLO
command (local lockout) and the
[Esc/Local] key does not work, switch the R&S NRT2 off and on again.
► Send the
>L
command (go to local).
► Tap the symbol on the touchscreen.
Chapter 4.1.2, "Status Information"
Remote Control