Measurement and Display Configuration
R&S
®
NRT2
41
User Manual 1178.5550.02 ─ 04
Standing Wave Ratio (SWR) ← Reflection
1 - Reflection coefficient
Standing wave ratio =
1 + Reflection coefficient
See also
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED
"POWer:SWRatio"
UNIT<Measurement>:POWer:REFLection
Return Loss ← Reflection
Reverse power
Return loss =
Forward power
10 x log
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED
"POWer:RLOSs"
UNIT<Measurement>:POWer:REFLection
Reflection Coefficient ← Reflection
Reflection coefficient =
Forward power
Reverse power
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED
"POWer:RCOefficient"
UNIT<Measurement>:POWer:REFLection
Reflection Ratio ← Reflection
Forward power
Reflection ratio =
Reverse power
100
Remote command:
CALCulate<Measurement>[:CHANnel<Channel>]:FEED
"POWer:RFRatio"
UNIT<Measurement>:POWer:REFLection
Fwd Unit
Sets the unit of the forward power measurement.
"dBm"
Power in dBm
"dB
μ
V"
Power in dB
μ
V
"W"
Power in W
Remote command:
UNIT<Measurement>:POWer[:VALue]
UNIT<Measurement>:POWer:RATio
Rel
Groups the settings for relative measurements.
Main Measurement Dialog