Remote Control
R&S
®
NRT2
156
User Manual 1178.5550.02 ─ 04
Example:
SENS2:FUNC:OFF "POW:REV"
SENS2: FUNC:OFF?
The following response would be possible:
"POW:CFAC","POW:FORW:AVER:BURS", "POW:FORW:PEP",
"POW:FORW:CCDF", "POW:ABS:AVER",
"POW:ABS:AVER:BURS", "POW:ABS:PEP", "POW:REV"
Usage:
Setting only
[SENSe<Sensor>:]FUNCtion:STATe?
<function>
This command returns the status of the measurement function:
●
Response 0: measurement function switched off
●
Response 1: measurement function switched on
All functions defined under
SENS<n>:FUNC:ON
are available as measurement func-
tions.
The command is a query and therefore has no *RST value.
Query parameters:
<function>
Example:
SENS2:FUNC:STAT? "POW:REV"
Usage:
Query only
[SENSe<Sensor>:]INFormation?
This command returns the condensed data of the sensor in form of an ASCII string.
Example:
SENS2:INF?
Usage:
Query only
Manual operation:
See
[SENSe<Sensor>:]NAME
<name>
Set a user selectable name for the sensor. This has no further influence on other func-
tions. It exists for naming a sensor according to user requirements.
Parameters:
<name>
[SENSe<Sensor>:]POWer:CCDFunction:REFerence
<ref>
This command defines the threshold for the cumulative distribution function (CCDF).
The distribution function states the probability (in %) of the envelope power lying above
the threshold. Units W or dB or dBm.
SENSe