Remote Control
R&S
®
NRT2
188
User Manual 1178.5550.02 ─ 04
Suffix:
<Measurement>
.
1...100
Parameters:
<unit>
DBM | DBUV | W
*RST:
DBM
Manual operation:
See
UNIT<Measurement>:POWer:REFLection
<unit>
This command defines the matching of the load being measured as standing wave
ratio, return loss, reflection coefficient, or power ratio R/F (in percent).
NOTE
: The command is only effective if the measurement function
POW:S11
or
POW:
REFL
has been activated, see SENS<n>:FUNC subsystem
Suffix:
<Measurement>
.
1...100
Parameters:
<unit>
RCO | RL | SWR | RFR
RCO
Reflection coefficient (0 to 1, without unit)
RL
Return loss (in dB)
SWR
Standing wave ratio (1 to ∞, without unit)
RFR
Ratio between forward and reverse power (0% to 100%)
*RST:
SWR
Example:
SENSe1:FUNC "POW:REFL"
UNIT1:POW:REFL RCO
Manual operation:
See
See
See
See
UNIT<Measurement>:POWer:RELative:STATe
<state>
This command defines the indication of the forward power (FWD), reverse power
(REV) and absorbed power (F-R) in absolute units (W, dBm) or in relative display
mode (%, dB).
Suffix:
<Measurement>
.
1...100
Parameters:
<state>
UNIT