Remote Control
R&S
®
NRT2
157
User Manual 1178.5550.02 ─ 04
Parameters:
<ref>
Range:
0.0 to 100.0e6
*RST:
1.0e-3
Default unit: W
Example:
SENS2:POW:CCDF:REF 10W
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
<state>
This command switches the automatic adaptation of the bargraph scaling for the reflec-
tion indication on or off.
Parameters:
<state>
ON
If auto-ranging is switched on, the scale limits of the bargraphs
are automatically adapted to the current measured value.
OFF
If it is switched off, the scale limits remain fixed.
*RST:
1
Example:
SENS1:POW:REFL:RANG:AUTO OFF
Manual operation:
See
See
[SENSe<Sensor>:]POWer:REFLection:RANGe:LIMit:DETect
<value>
This command defines the conditions for a logic high level (> 2.7 V) being output at the
AUX TTL connector if this connector has been defined as a monitoring output for the
power indication (e.g. with the command
SENSe<n>:POWer:RANGe:LIMit ON
).
Parameters:
<value>
INBound | OUTBound | HIGH
INBound
A high level is output if the measured power is within the range
specified by the scale limits of the lefthand bargraph.
OUTBound
A high level is output if the measured power is out of the range
defined by the scale limits of the lefthand bargraph.
HIGH
A high level is output if the measured power exceeds the upper
scale limit of the lefthand bargraph.
*RST:
HIGH
Example:
SENS1:POW:RANG:LIM:DET INB
Manual operation:
See
SENSe