1.1 Methods
Notes:
1. Run samples and blanks with the same batch of vials. The blank is stable when stored in
the dark and can be used for further measurements with vials of the same batch.
2. Do not place the hot vials in the sample chamber. Cool the vials to room temperature for
final measurements.
3. Suspended solids in the vial lead to incorrect measurements. For this reason it is
important to place the vials carefully in the sample chamber. The precipitate at the
bottom of the sample should be not suspended.
4. Clean the outside of the vials with a towel. Finger prints or other marks will be removed.
5. Samples can be measured when the Chloride contents does not exceed 1000 mg/l.
6. In exceptional cases, compounds contained in the water cannot be oxidized adequate,
what results in minimum findings, compared with the reference method.
7. For samples under 100 mg/l COD it is recommendable to repeat the test with the tube
test for COD LR.
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...