16
SP600 Spectrophotometer 01/2008
1.1 Methods
Alkalinity-p = p-value
with Tablet
5 – 300 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with
10 ml of the water sample
,
close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
one ALKA-P-PHOTOMETER tablet
straight
from the foil to the water sample and crush the tablet
using a clean stirring rod.
6. Close the vial tightly with the cap and swirl several
times until the tablet is dissolved.
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press
TEST
key.
The result is shown in the display as Alkalinity-p.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
3
5
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...