291
SP600 Spectrophotometer 01/2008
Delete stored results
Press [MODE] [3] [4] keys.
Confirm with
[ ]
key.
The display shows:
• Press [0] key to retain the data sets in memory.
• After pressing key [1] the following acknowledgment is
displayed:
Press
[ ]
key to delete.
ATTENTION:
All stored test results are deleted
or cancel without deleting data by pressing [ESC] key.
Note:
All stored test results are deleted.
2.4.5 Calibration
Calibration Fluoride
Regard notes!
Press [MODE] [4] [0] keys.
Confirm with [ ] key.
The display shows:
1. Fill a clean vial (24 mm Ø) with exact
10 ml of deionized
water
, close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
<Delete data>
Delete all data?
YES : 1 NO : 0
<Calibration>
170 Fluoride
Zero: deionised water
press ZERO
<Delete data>
Delete data
Do not delete: ESC
Mode
3
4
1
0
Mode
4
0
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...