ORBECO SP600 Instruction Manual Download Page 199

197

SP600 Spectrophotometer  01/2008

1.1 Methods

Notes:

1. Vial cleaning:
  As many household cleaners (e.g. dishwasher detergent) contain reducing  

 

  substances, the subsequent determination of Ozone may show lower results. To avoid  
  any measurement errors, only use glassware free of Chlorine consumption.
  Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for  
  one hour, then rinse all glassware thoroughly with deionized water. 
2. Preparing the sample:
  When preparing the sample, the escape of Ozone gases, e.g. by pipetting or shaking,  
  must be avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.3 to 6.5. The reagent  
  tablet  therefore contains a buffer for the pH adjustment.
  Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7  
  before the tablet is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Turbidity (lead to errors):
  The use of the DPD No. 1 tablet in samples with high Calcium ion contents* and/or high  
  conductivity* can lead to turbidity of the sample and therefore incorrect measurements.
    

* it is not possible to give exactly values, because the development of    

 

  turbidity depends on nature and ingredients of the sample.

 

5. Exceeding of the measuring range:
  Concentrations above 5 mg/l Ozone can produce results within the measuring range  
  up to 0 mg/l. In this event, the water sample must be diluted with water free of Ozone.  
  10 ml of the diluted sample will be mixed with the reagent and the measurement  
  repeated.
6. If   ???   is displayed at the diffentiated test result see page 325.

Oxidizing agents such as Bromine, Chlorine etc. interfere as they react like Ozone.

Summary of Contents for SP600

Page 1: ...Instruction Manual SP600 Spectrophotometer ...

Page 2: ......

Page 3: ...Be aware of the ingredients and take proper care when disposing of the test solution CAUTION Please read this instruction manual before unpacking setting up or using the photometer Please read the method description completely before performing the test Be aware of the risks of using the required reagents by reading the MSDS Material Safety Data Sheets Failure could result in serious injury to the...

Page 4: ...Chlorine 50 Chlorine with Tablet differentiated determination free combined total 52 55 58 free Chlorine 53 56 60 total Chlorine 54 57 61 Chlorine with Liquid Reagent differentiated determination free combined total 62 free Chlorine 64 total Chlorine 65 Chlorine Powder Pack differentiated determination free combined total 66 free Chlorine 68 total Chlorine 69 Chlorine HR Kl 70 Chlorine dioxide in ...

Page 5: ...total HR 128 Hazen 130 Hydrazine 132 Hydrazine with Liquid Reagent 134 Hydrogen peroxide 136 138 Iodine 140 Iron 142 Iron with Tablets 144 146 148 Iron Powder Packs 150 Iron TPTZ Powder Packs 152 Lead 154 Lead Procedure A 156 158 Lead Procedure B 156 159 Mangenese with Tablet 160 Manganese LR Powder Packs 162 Manganese HR Powder Packs 164 Molybdate with Tablets 166 Molybdate HR Powder Packs 168 Ni...

Page 6: ...ablets 220 Phosphate ortho HR with Tablets 222 Phosphate ortho Powder Packs 224 Phosphate ortho Tube Test 226 Phosphate ortho Tube Test 228 Phosphate hydrolysable Tube Test 230 Potassium 232 S Abs Spectral Absorption Coefficient 234 Silica 236 Silica LR PP 238 Silica HR PP 240 Sulfate Powder Pack 242 Sulfide 244 Sulfite 246 248 Surfactants anionic 250 TOC HR 252 Turbidity 254 Urea 256 Zinc 258 1 2...

Page 7: ...fferentiation 272 2 3 4 Performing Zero 272 2 3 5 Performing Test 273 2 3 6 Ensuring reaction periods countdown 273 2 3 7 Changing chemical species 274 2 3 8 Storing results 274 2 3 9 Printing results 275 2 3 10 Perform additional measurements 276 2 3 11 Selecting a new method 276 2 4 Photometer settings Mode Menu 277 2 4 1 blank because of technical requirements 2 4 2 Instrument basic settings 1 ...

Page 8: ...tent 321 3 3 blank because of technical requirements 3 4 Technical data 322 3 5 Abbreviations 323 3 6 Troubleshooting 324 3 6 1 Operating messages in the display error display 324 3 6 2 General problems 326 3 6 3 Service Maintenance 327 3 6 3 1 Handling Cleaning 327 3 6 3 2 Changing the light source 327 3 6 3 3 blank because of technical requirements 3 6 3 4 Changing the batteries 329 3 7 Declarat...

Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...

Page 10: ... mg l N Salicylate2 655 30 68 Arsenic see test instructions 0 02 0 6 mg l As Silverdiethyldithio carbamate1 507 32 85 Boron T tablet 0 1 2 mg l B Azomethine3 450 36 78 Bromine 10 T tablet 0 1 3 mg l Br2 DPD5 510 38 79 Bromine 50 T tablet 0 05 1 mg l Br2 DPD5 510 40 80 Bromine T tablet 0 05 6 5 mg l Br2 DPD5 510 42 87 Cadmium TT tube test 0 025 0 75 mg l Cd Cadion6 525 44 90 Chloride T tablet 0 5 2...

Page 11: ...acid 1 585 110 160 Cyanuric acid T tablet 2 160 mg l Cys Melamine 530 112 165 DEHA T tablet liquid 20 500 μg l DEHA PPST3 562 114 167 DEHA PP powder liquid 20 500 μg l DEHA PPST3 562 116 170 Fluoride L liquid 0 05 1 5 mg l F SPADNS2 580 118 175 Formaldehyde 10 powder liquid 1 5 mg l HCHO H2 O2 Chromotropic acid6 585 120 176 Formaldehyde 50 powder liquid 0 02 1 mg l HCHO H2 O2 Chromotropic acid6 58...

Page 12: ... 1 5 mg l Pb 4 2 Pyridylazo resorcin6 515 156 159 240 Manganese T tablet 0 2 4 mg l Mn Formaldoxime 450 160 242 Manganese LR PP powder pack liquid 0 01 0 7 mg l Mn PAN 558 162 243 Manganese HR PP powder pack 0 1 18 mg l Mn Periodate oxidation2 525 164 250 Molybdate T tablet 1 30 mg l MoO4 Thioglycolate4 366 166 252 Molybdate HR PP powder pack 0 5 66 mg l MoO4 Mercaptoacetic acid 420 168 255 Nickel...

Page 13: ...tablet 6 5 8 4 Phenol red5 558 206 331 pH Value L liquid 6 5 8 4 Phenol red5 558 208 315 Phenol T tablet 0 1 5 mg l C6 H5 OH 4 Aminoantipyrine 1 507 210 326 Phosphate total TT tube test 0 02 1 1 mg l P Acid persulf digestion Ascorbic acid2 890 212 214 317 Phosphate total LR TT tube test 0 07 3 mg l P Phosphomolybdic acid Ascorbic acid2 690 212 216 318 Phosphate total HR TT tube test 1 5 20 mg l P ...

Page 14: ...238 352 Silica HR PP powder pack 1 100 mg l SiO2 Silicomolybdate 452 240 360 Sulfate PP powder pack 2 100 mg l SO4 Bariumsulfate Turbidity2 450 242 365 Sulfide tablet 0 04 0 5 mg l S DPD Catalyst3 4 668 244 368 Sulfite 10 T tablet 0 1 10 mg l SO3 DTNB 405 246 370 Sulfite T tablet 0 05 4 mg l SO3 DTNB 405 248 375 Surfactants anionic TT tube test 0 05 2 mg l MBAS Methylene blue6 1 653 250 381 TOC HR...

Page 15: ...ftliche Verlagsgesellschaft mbH Stuttgart 1989 4 Photometrische Analyse Lange Vejdelek Verlag Chemie 1980 5 Colorimetric Chemical Analytical methods 9th Edition London 6 adapted from Merck for more information see instructions delivered with the test Notes for searching Active Oxygen Oxygen activ Alkalinity m Alkalinity total Alkalinity total Alkalinity total Colour Hazen or Spectral Absorption Co...

Page 16: ...e the vial from the sample chamber 5 Add one ALKA M PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as Acid demand to...

Page 17: ...pectrophotometer 01 2008 1 1 Methods Notes 1 The terms total Alkalinity Alkalinity m m Value and Acid demand to pH 4 3 are identical 2 For accurate results exactly 10 ml of water sample must be taken for the test ...

Page 18: ...key 4 Remove the vial from the sample chamber 5 Add one ALKA P PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as Alk...

Page 19: ...3 0 80 1 43 CaCO3 dH eH fH aH 5 By determining Alkalinity p and Alkalinity m it is possible to classify the alkalinity as Hydroxide Carbonate and Hydrogencarbonate The following differentiation is only valid if a no other alkalis are present and b Hydroxide und Hydrogen are not present in the same water sample If condition b is not fulfilled please get additional information from Deutsche Einheits...

Page 20: ...s ZERO key 4 Remove the vial from the sample chamber 5 Add one ALKA M PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display...

Page 21: ...ate results exactly 10 ml of water sample must be taken for the test 3 Conversion table Acid demand to pH 4 3 DIN 38 409 KS4 3 German dH English eH French fH 1 mg l CaCO3 0 02 0 056 0 07 0 1 Carbonate hardness reference Hydrogencarbonate anions Example 10 mg l CaCO3 10 mg l x 0 056 0 56 mg l dH 10 mg l CaCO3 10 mg l x 0 02 0 2 mmol l 4 CaCO3 dH eH fH aH ...

Page 22: ... clean stirring rod dissolve the tablet 6 Add one ALUMINIUM No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl gently several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 5 minu...

Page 23: ...y insignificant unless the water has fluoride added artificially In this case the following table should be used Fluoride mg l F Displayed value Aluminium mg l Al 0 05 0 10 0 15 0 20 0 25 0 30 0 2 0 05 0 11 0 16 0 21 0 27 0 32 0 4 0 06 0 11 0 17 0 23 0 28 0 34 0 6 0 06 0 12 0 18 0 24 0 30 0 37 0 8 0 06 0 13 0 20 0 26 0 32 0 40 1 0 0 07 0 13 0 21 0 28 0 36 0 45 1 5 0 09 0 20 0 29 0 37 0 48 Example ...

Page 24: ...period is finished proceed as follows 5 Add one Vario Hexamine F20 powder pack straight from the foil to the same water sample 6 Dissolve the powder using a clean stirring rod 7 Add 1 drop of Vario Aluminum ECR Masking Re agent in the vial marked as blank 8 Add 10 ml of the prepared water sample to the vial this is the blank 9 Add the remaining 10 ml of the prepared water sample in the second clea...

Page 25: ...mperature must be between 20 C and 25 C 3 A low test result may be given in the presence of Fluorides and Polyphosphates The effect of this is generally insignificant unless the water has fluoride added artificially In this case the following table should be used Fluoride mg l F Displayed value Aluminium mg l Al 0 05 0 10 0 15 0 20 0 25 0 30 0 2 0 05 0 11 0 16 0 21 0 27 0 32 0 4 0 06 0 11 0 17 0 2...

Page 26: ...blet using a clean stirring rod 6 Add one AMMONIA No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the re...

Page 27: ...our development At a temperature below 20 C the reaction period is 15 minutes 4 Sea water samples Ammonia conditioning reagent is required when testing sea water or brackish water samples to prevent precipitations of salts Fill the test tube with the sample to the 10 ml mark and add one level spoonful of Conditioning Powder Mix to dissolve then continue as described in the test instructions 5 Conv...

Page 28: ...n period is finished proceed as follows 6 Add one Vario Ammonium Cyanurate F10 powder pack straight from the foil to each sample 7 Close the vials with the caps tightly and shake to mix the contents 8 Press key Wait for a reaction period of 15 minutes After reaction period is finished proceed as follows 9 Place the vial the blank in the sample chamber making sure that the marks are aligned 10 Pres...

Page 29: ...forming a total Iron test b add the same iron concentration as determined to the deionized water step 1 The interference will be blanked out successfully Magnesium greater than 6000 mg l CaCO3 Nitrate greater than 100 mg l NO3 N Nitrite greater than 12 mg l NO2 N Phosphate greater than 100 mg l PO4 P Sulfate greater than 300 mg l SO4 Sulfide intensifies the colour Glycine Hydrazine Colour Turbidit...

Page 30: ...m the foil into each vial 5 Close the vials tightly and swirl several times to dissolve the powder 6 Press key Wait for a reaction period of 20 minutes After reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place the vial the sample in the sample chamb...

Page 31: ... add one drop of 0 1 mol l Sodium thiosulfate for each 0 3 mg l Cl2 in a one litre water sample 3 Iron interferes with the test The interferences will be eliminated as follows Determine the amount of total iron present in the water sample To produce the blank add an iron standard solution with the same iron concentration to the vial point 1 instead of deionized water 4 Conversion mg l NH4 mg l N x...

Page 32: ...m the foil into each vial 5 Close the vials tightly and swirl several times to dissolve the powder 6 Press key Wait for a reaction period of 20 minutes After reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place the vial the sample in the sample chamb...

Page 33: ...ent add one drop of 0 1 mol l Sodium thiosulfate for each 0 3 mg l Cl2 in a one litre water sample 3 Iron interferes with the test The interferences will be eliminated as follows Determine the amount of total iron present in the water sample Add an iron standard solution with the same concentration to the vial point 1 instead of deionized water to produce the blank 4 Conversion mg l NH4 mg l N x 1...

Page 34: ...5 g Silver diethyldithiocarbamate C5 H10 AgNS2 p a and 0 02 g Brucine C23 H26 N2 O4 p a in 100 ml 1 Methyl 2 pyrrolidone extra pure C5 H9 NO and store in a dark bottle If it is not possible to dissolve completely stir for min 1 hour and filtrate to get a clear solution Notes use only dry glass vessels stored in a dark glass bottle at max 20 C the absorp tion solution can be used for about 1 week s...

Page 35: ...he 15 minutes reaction time add the 2 g Zinc to the Erlenmeyer flask and immediately as semble the apparatus with the prepared absorption tube see picture 8 The reaction starts fume hood Wait for 60 minutes reaction time Performing test procedure 9 Fill a clean 20 mm cell note 1 with deionized water 10 Place the cell in the sample chamber making sure that the positioning is correct 11 Press ZERO k...

Page 36: ...ered locally MSDS please refer to your local reagent supplier Ensure proper disposal of reagent solution 3 Use a cell with 20 mm path length Order code 60 10 50 Positioning insert cell on the left side in the sample chamber c clip 4 According to literature G Ackermann J Köthe Fresenius Z Anal Chem 323 1986 135 Sb Se and Te interfere due to the same reaction Thiosulfate interferes differently ...

Page 37: ...35 SP600 Spectrophotometer 01 2008 ...

Page 38: ... clean stirring rod and dissolve the tablet 6 Add one BORON No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 20 minutes Af...

Page 39: ...ded in the correct sequence 2 The sample solution should have a pH value between 6 and 7 3 Interferences are prevented by the presence of EDTA in the tablets 4 The rate of colour development depends on the temperature The temperature of the sample must be 20 C 1 C 5 B H3 BO3 ...

Page 40: ...e a beaker with the water sample and empty it leaving a few drops remaining in the beaker 6 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 10 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The res...

Page 41: ...g must be avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding the...

Page 42: ...se a beaker with the water sample and empty it leaving a few drops remaining in the beaker 6 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The re...

Page 43: ...g must be avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding the...

Page 44: ...e sample chamber empty the vial leaving a few drops in 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in t...

Page 45: ...be avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding of the...

Page 46: ...mix the contents 6 Add 0 2 ml reagent Cd 1K 7 Close the vial tightly with the cap and invert several times to mix the contents 8 Add one level microspoon of reagent Cd 2K 9 Close the vial tightly with the cap and swirl until the reagent is solved completely 10 Place the vial in the sample chamber making sure that the marks are aligned 11 Press TEST key Wait for a reaction period of 2 minutes After...

Page 47: ...pany MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature the sample temperature must be between 10 and 40 C 6 Sample and reagent volumes should always metered by using volumetric pipettes class A 7 This test determines only Cd2 ions Samples must be pre treated or decomposed by digestion before colloi...

Page 48: ...stirring rod and dissolve the tablet 6 Add one CHLORIDE T2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl gently several times until the tablet is dissolved Note 1 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minut...

Page 49: ...eads to bigger sized particles which can cause false readings 2 High concentrations of electrolytes and organic compounds have different effects on the precipitation reaction 3 Ions which also form deposits with Silver nitrate in acidic media such as Bromides Iodides and Thiocyanates interfere with the analysis 4 Highly alkaline water should if necessary be neutralised using Nitric acid before ana...

Page 50: ...rt the vials several times to mix the contents 5 Fill each vial with the same size drops by holding the bottle vertically and squeeze slowly 3 drops Chlorid 52 6 Close the vials with the caps tightly and invert the vials several times to mix the contents 7 Press key Wait for a reaction period of 3 minutes 8 Place the vial the blank in the sample chamber making sure that the marks are aligned 9 Pre...

Page 51: ...hould have room temperature for test performance 2 The test sample should have a pH of between 3 and 9 3 Store the reagent bottles in a cool dry place ideally at between 4 C and 8 C 4 Interferences Thiocyanate Sulfide Thiosulfate Bromide and Iodide interfere because they react like Chloride ...

Page 52: ... 3 mg l Cl2 Chlorine with Powder Pack 0 01 2 mg l Cl2 The following selection is shown in the display for the differentiated determination of free combined and total Chlorine for the determination of free Chlorine for the determination of total Chlorine Select the desired determination with the arrow keys and Confirm with key Chlorine diff free total diff free total 9 8 9 9 1 0 0 1 0 1 1 1 0 ...

Page 53: ...efore the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 5 Exceeding of the measuring range Concentrations above 10 mg l Chlorine using tablets 4 mg l Chlorine using liquid reagents 2 mg l using powder packs can produce results within the measuring range up to 0 mg l In this event the water sample must be diluted with water free of Chlorine 10 ml of the diluted sample w...

Page 54: ... sample chamber making sure that the positioning is correct 10 Press TEST key 11 Remove the cell from the sample chamber and return the colored test solution completely back into the bea ker 12 Add one DPD No 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod Dissolve the tablet 13 Fill the 10 mm cell with the colored test solution 14 Place the...

Page 55: ...r with the water sample and empty it leaving a few drops remaining in the beaker 6 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 10 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The result is sh...

Page 56: ... beaker 6 Add one DPD No 1 and one DPD No 3 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 10 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is fin...

Page 57: ...e sample chamber making sure that the positioning is correct 10 Press TEST key 11 Remove the cell from the sample chamber and return the colored test solution completely back into the bea ker 12 Add one DPD No 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod Dissolve the tablet 13 Fill the 50 mm cell with the colored test solution 14 Place th...

Page 58: ...ker with the water sample and empty it leaving a few drops remaining in the beaker 6 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The result is ...

Page 59: ...ing in the beaker 6 Add one DPD No 1 and one DPD No 3 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 2 minutes After the reaction per...

Page 60: ...from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key 10 Remove the vial from the sample chamber 11 Add one DPD No 3 tablet straight from the foil to the same water sam...

Page 61: ...ss TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes see page 51 T1 accepted prepare T2 press TEST Countdown 2 00 mg l free Cl mg l comb Cl mg l total Cl ...

Page 62: ...amber empty the vial leaving a few drops in 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the display ...

Page 63: ...DPD No 3 tablet straight from the foil and crush the tablets using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the meas...

Page 64: ... size drops by holding the bottle vertically and squeeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key 10 Remove the vial from the sample chamber 11 Add 3 dro...

Page 65: ... is finished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes 1 After use replace the bottle caps securely noting the colour coding 2 Store the reagent bottles in a cool dry place ideally between 6 C and 10 C 3 Also see page 51 T1 accepted prepare T2 press TEST Countdown 2 00 mg l free Cl mg l comb Cl mg ...

Page 66: ... squeeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the display in mg l free Chlorine Notes free and total Chlorine 1 After use repl...

Page 67: ... vertically and squeeze slowly 6 drops of DPD 1 buffer solution 2 drops of DPD 1 reagent solution 3 drops of DPD 3 solution 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction per...

Page 68: ...he water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key 9 Remove the vial from the sample chamber Empty the vial rinse vial and cap several times and then fill the vial with 10 ml of water sample 10 Add one VARIO Chlorine TOTAL DPD F10 powder pa...

Page 69: ...ss TEST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l free Chlorine mg l combined Chlorine mg l total Chlorine Notes see page 51 mg l free Cl mg l comb Cl mg l total Cl T1 accepted prepare T2 press TEST Countdown 3 00 ...

Page 70: ...Remove the vial from the sample chamber 5 Add one VARIO Chlorine FREE DPD F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l free Chlorine Notes see page ...

Page 71: ...TOTAL DPD F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several time to mix the contents approx 20 seconds 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automatically The result is shown in...

Page 72: ... straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Add one ACIDIFYING GP tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press T...

Page 73: ...71 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidizing agents interfere as they react like Chlorine ...

Page 74: ...and dry completely 5 Rinse a beaker with the sample and empty it leaving a few drops in 6 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The resul...

Page 75: ...e pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Turbidity lead to errors The use of the DPD No 1 tablet in samples with high Calcium ion contents and or high conductivity can lead to turbidity of the sample and therefore incorrect measurements In this event the reag...

Page 76: ...g selection is shown in the display for the determination of Chlorine dioxide in the presence of Chlorine for the determination of Chlorine dioxide in the absence of Chlorine Select the desired determination with the arrow keys and Confirm with key Chlorine dioxide with Cl without Cl with Cl without Cl 1 2 0 ...

Page 77: ...lace immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding of the measuring range Concentrations abov...

Page 78: ...om the foil and crush the tablet using a clean stirring rod 6 Fill a second clean vial with 10 ml of water sample 7 Add one GLYCINE tablet straight from the foil and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 Transfer the contents of the second vial into the prepared vial 10 Close the vial tightly with t...

Page 79: ...d one DPD No 3 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 21 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 22 Place the vial in the sample chamber making sure that the marks are aligned 23 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measuremen...

Page 80: ...ly 0 38 mg l ClO2 mg l ClO2 Cl x 0 38 ClO2 Cl ClO2 Chlorine dioxide displayed as Chlorine units ClO2 Cl has its origin out of the swimming poolwater treatment according to DIN 19643 2 The total Chlorine result given includes the contribution by the Chlorine dioxide as Chlorine reading For true total Chlorine value subtract the Chlorine dioxide as Chlorine reading from the quoted total Chlorine rea...

Page 81: ...eaving a few drops in 5 Add one DPD No 1 tablet straight from the foil and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is shown in the display as Chlorine dioxide in...

Page 82: ...owing selection is shown in the display for the differentiated determination of Chromium VI Chromium III and total Chromium for the determination of Chromium VI for the determination of total Chromium sum Cr III Cr VI Select the desired determination with the arrow keys and Confirm with the key Chrom diff Cr VI Cr III VI diff Cr VI Cr III VI 1 2 4 1 2 5 ...

Page 83: ...81 SP600 Spectrophotometer 01 2008 ...

Page 84: ...g test procedure 6 Fill a clean 50 mm cell with deionized water 7 Place the cell in the sample chamber making sure that the positioning is correct 8 Press ZERO key 9 Remove the cell from the sample chamber Empty the cell and dry completely 10 Add one CHROMIUM HEXAVALENT powder pack straight from the foil into the pre prepared vial see step 5 11 Close the vial tightly with the cap and swirl several...

Page 85: ...inished the measurement starts automatically The result is shown in the display in mg l Cr Vl mg l Cr lll mg l Cr total chromium Notes 1 Performing steps 1 14 determines concentration of total chromium and steps 15 20 determines concentration of Chromium VI The concentration of Chromium III results out of the difference 2 pH value of the water sample should be between 3 and 9 3 For information abo...

Page 86: ...dd one CHROMIUM HEXAVALENT powder pack straight from the foil to the water sample 7 Close the vial tightly with the cap and swirl several times to mix the contents 8 Fill the 50 mm cell with this test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measure...

Page 87: ...ean 50 mm cell with deionized water 7 Place the cell in the sample chamber making sure that the positioning is correct 8 Press ZERO key 9 Remove the cell from the sample chamber Empty the cell and dry completely 10 Add one CHROMIUM HEXAVALENT powder pack straight from the foil to the pre prepared water sample 11 Close the vial tightly with the cap and swirl several times to mix the contents 12 Fil...

Page 88: ...nvert the vial and allow to cool to room temperature Performing test procedure 6 Place the pre prepared vial in the sample chamber making sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Add one CHROMIUM HEXAVALENT powder pack straight from the foil into the pre prepared vial 10 Close the vial tightly with the cap and swirl several times to mix the conte...

Page 89: ...ent starts automatically The result is shown in the display in mg l Cr Vl mg l Cr lll mg l Cr total Chromium Notes 1 Performing steps 1 12 determines concentration of total chromium and steps 13 17 determines concentration of Chromium VI The concentration of Chromium III results out of the difference 2 pH value of the water sample should be between 3 and 9 3 For information about interferences esp...

Page 90: ... powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l Chromium ...

Page 91: ... to room temperature Performing test procedure 6 Place the pre prepared vial in the sample chamber ma king sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Add one CHROMIUM HEXAVALENT powder pack straight from the foil to the water sample 10 Close the vial tightly with the cap and swirl several times to mix the contents 11 Place the vial in the sample ch...

Page 92: ...perature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are aligned 7 Press ZERO key 8 Re...

Page 93: ... Suspended solids in the vial lead to incorrect measurements For this reason it is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride contents does not exceed 1000 mg l 6 In exceptional c...

Page 94: ...mperature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are aligned 7 Press ZERO key 8 R...

Page 95: ...surements For this reason it is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride contents does not exceed 1000 mg l 6 In exceptional cases compounds contained in the water cannot be oxi...

Page 96: ...r at a temperature of 150 C 5 CAUTION The vials are hot Remove the tubes from the heating block and allow them to cool to 60 C or less Mix the contents by carefully inverting each tube several times while still warm Then allow the tubes to cool to ambient temperature before measuring Note 2 6 Place the vial the blank Note 3 4 in the sample chamber making sure that the marks are aligned 7 Press ZER...

Page 97: ... is important to place the vials carefully in the sample chamber The precipitate at the bottom of the sample should be not suspended 4 Clean the outside of the vials with a towel Finger prints or other marks will be removed 5 Samples can be measured when the Chloride contents does not exceed 1000 mg l 6 In exceptional cases compounds contained in the water cannot be oxidized adequate what results ...

Page 98: ...he following selection is shown in the display for the differentiated determination of free combined and total Copper for the determination of free Copper for the determination of total Copper Select the desired determination with the arrow keys and Confirm with key Copper diff free total diff free total 1 4 9 1 5 0 ...

Page 99: ...97 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If is displayed at the diffentiated test result see page 325 ...

Page 100: ...OPPER No 1 tablet straight from the foil and crush the tablet using a clean stirring rod Dissolve the tablet 7 Fill the 50 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key 10 Remove the cell from the sample chamber and return the colored test solution completely back into the beaker 12 Add one COPPER No 2 tab...

Page 101: ...red test solution 13 Place the cell in the sample chamber making sure that the positioning is correct 14 Press TEST key The result is shown in the display in mg l free Copper mg l combined Copper mg l total Copper mg l free Cu mg l comb Cu mg l tot Cu T1 accepted prepare T2 press TEST ...

Page 102: ...ber Empty the cell and dry completely 5 Fill a beaker with 10 ml of water sample 6 Add one COPPER No 1 tablet straight from the foil and crush the tablet using a clean stirring rod Dissolve the tablet 7 Fill the 50 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key The result is shown in the display in mg l Cop...

Page 103: ...l and dry completely 5 Fill a beaker with 10 ml of water sample 6 Add one COPPER No 1 tablet and one COPPER No 2 tablet straight from the foil and crush the tablets using a clean stirring rod Dissolve the tablets 7 Fill the 50 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key The result is shown in the display...

Page 104: ...il to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key 9 Remove the vial from the sample chamber 10 Add one COPPER No 2 tablet straight from the foil to the same water sample and crush the tablet u...

Page 105: ...e the vial in the sample chamber making sure that the marks are aligned 13 Press TEST key The result is shown in the display in mg l free Copper mg l combined Copper mg l total Copper mg l free Cu mg l comb Cu mg l total Cu T1 accepted prepare T2 press TEST ...

Page 106: ... Remove the vial from the sample chamber 5 Add one COPPER No 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l free Copp...

Page 107: ...l from the sample chamber 5 Add one COPPER No 1 tablet and one COPPER No 2 tablet straight from the foil to the water sample and crush the tablets using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in...

Page 108: ...r 5 Add one VARIO Cu 1 F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents Note 3 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown...

Page 109: ...ferences Cyanide CN Cyanide prevents full colour development Add 0 2 ml Formaldehyde to 10 ml of water sample and wait for a reaction time of 4 minutes Cyanide is masked After this perform test as described Multiply the result by 1 02 to correct the sample dilution by Formaldehyde Silver Ag If a turbidity remains and turns black silver interferences is likely Add 10 drops of saturated Potassium ch...

Page 110: ...anide 11 into the prepared water sample Dissolve reagent 7 Add 2 level spoons of No 4 grey Cyanide 12 and dissolve reagent 8 Fill the cell with the same size drops by holding the bottle vertically and squeeze slowly 3 drops of Cyanide 13 9 Fill the 50 mm cell with the colored test solution 10 Place the cell in the sample chamber making sure that the positioning is correct 11 Press TEST key Wait fo...

Page 111: ... destroyed by Chlorine are determined by this test 2 In the present of Thiocyanate heavy metal complexes colorants or aromatic amines the cyanide must be separated out by distillation before analysis is performed 3 Store the reagents in closed containers at a temperature of 15 C to 25 C 4 mg l µg l ...

Page 112: ... contents 6 Add 2 level spoons of No 4 grey Cyanide 12 clo se the vial tightly with the cap and invert the vial several times to mix the contents 7 Fill the vial with the same size drops by holding the bottle vertically and squeeze slowly 3 drops of Cyanide 13 8 Close the vial tightly with the cap and invert the vial several times to mix the contents 9 Place the vial in the sample chamber making s...

Page 113: ...can be destroyed by Chlorine are determined by this test 2 In the present of Thiocyanate heavy metal complexes colorants or aromatic amines the cyanide must be separated out by distillation before analysis is performed 3 Store the reagents in closed containers at a temperature of 15 C to 25 C ...

Page 114: ...s ZERO key 4 Remove the vial from the sample chamber 5 Add one CYANURIC ACID tablet straight from the foil to the prepared water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved Note 2 3 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown ...

Page 115: ...ater free of Cyanuric acid 2 If Cyanuric acid is present a cloudy solution will be given Single particles are uncaused necessarily by Cyanuric acid 3 Dissolve the tablet completely therefore swirl the vial approx 1 minute Not dissolved particles of the tablet can cause too high results ...

Page 116: ... 6 drops 0 25 ml of DEHA reagent 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Add one DEHA tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 Place the vial in the sample chamber making sure that the marks are align...

Page 117: ...Repeat the test procedure but without adding the DEHA solution If the displayed result is above 20 μg l subtract this value from the DEHA test result Substances which reduce Iron III interfere Substances which complex iron strongly may interfere also Substances which may interfere when present in concentrations at Borate as Na2 B4 O7 500 mg l Cobalt 0 025 mg l Copper 8 0 mg l Hardness as CaCO3 100...

Page 118: ...eral times to mix the contents 5 Add 0 20 ml VARIO DEHA 2 Rgt Solution to each vial Note 4 6 Close the vials tightly and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes Note 5 After reaction period is finished proceed as follows 8 Place the vial the blank in the sample chamber making sure that the marks are aligned 9 Press ZERO key 10 Remove the vial fr...

Page 119: ...interferes at all concentrations Repeat the test procedure but without adding the VARIO DEHA Rgt 2 solution If the displayed result is above 20 μg l subtract this value from the DEHA test result Substances which reduce Iron III interfere Substances which complex iron strongly may interfere also Substances who may interfere when present in concentrations at Borate as Na2 B4 O7 500 mg l Cobalt 0 025...

Page 120: ...d 3 Press ZERO key 4 Remove the vial from the sample chamber 5 Add exactly 2 ml SPADNS reagent solution Note 4 to the water sample Caution Vial is filled up to the top Note 8 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l Fluor...

Page 121: ...amples to be tested should have the same temperature 1 C 4 As the test result is highly dependent on exact sample and reagent volumes the sample and reagent volumes should always be metered by using a 10 ml resp 2 ml volumetric pipette class A 5 The accuracy of the test methods decreases above a level of 1 2 mg l Fluoride Although the results are sufficiently accurate for most applications even mo...

Page 122: ...ith 3 ml of the water sample this is the test solution 6 Close the vials tightly and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes After the reaction period is finished proceed as follows 8 Fill the 10 mm cell with the prepared blank 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press ZERO key 11 Remove the cell...

Page 123: ...t and the MSDS available at www merck com 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature sample and tube temperature must be between 20 and 25 C 6 Sample volume should always metered by using volumetric pipette class A ...

Page 124: ...with 3 ml of the water sample this is the test solution 6 Close the vials tightly and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes After the reaction period is finished proceed as follows 8 Fill the 50 mm cell with the prepared blank 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press ZERO key 11 Remove the cel...

Page 125: ...t and the MSDS available at www merck com 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature sample and tube temperature must be between 20 and 25 C 6 Sample volume should always metered by using volumetric pipette class A ...

Page 126: ...vigorously until the reagent is completely dissolved 6 Add 2 ml of water sample Note 6 7 Close the vial tightly with the cap Hold the vial by the cap and carefully invert several times to mix the contents CAUTION Vial becomes hot 8 Press key Wait for a reaction period of 5 minutes After reaction period is finished proceed as follows 9 Place the vial in the sample chamber making sure that the marks...

Page 127: ...t and the MSDS available at www merck com 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature sample and tube temperature must be between 20 and 25 C 6 Sample volume should always metered by using volumetric pipette class A ...

Page 128: ...ablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl gently several times to dissolve the tablet 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically T...

Page 129: ...ter samples must be adjusted between pH 4 and pH 10 before the tablet is added use 1 mol l Hydrochloric acid resp 1mol l Sodium hydroxide 2 Conversion table mg l CaCO3 dH fH eH 1 mg l CaCO3 0 056 0 10 0 07 1 dH 17 8 1 78 1 25 1 fH 10 0 0 56 0 70 1 eH 14 3 0 80 1 43 3 CaCO3 dH eH fH aH ...

Page 130: ...5 Add one HARDCHECK P tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl gently several times to dissolve the tablet 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement...

Page 131: ...ter samples must be adjusted between pH 4 and pH 10 before the tablet is added use 1 mol l Hydrochloric acid resp 1mol l Sodium hydroxide 2 Conversion table mg l CaCO3 dH fH eH 1 mg l CaCO3 0 056 0 10 0 07 1 dH 17 8 1 78 1 25 1 fH 10 0 0 56 0 70 1 eH 14 3 0 80 1 43 3 CaCO3 dH eH fH aH ...

Page 132: ...h deionized water 3 Place the cell in the sample chamber making sure that the positioning is correct 4 Press ZERO key 5 Remove the cell from the sample chamber Empty the cell 6 Rinse the 50 mm cell with the water sample and fill with water sample 7 Place the cell in the sample chamber making sure that the positioning is correct 8 Press TEST key The result is shown in the display in mg l Pt Co unit...

Page 133: ...en is calibrated on the basis of the standards specified by Standard Methods for the Examination of Water and Wastewater also see EN ISO 7887 1994 1 Pt Co colour unit 1 mg L of platinum as chloroplatinate ion 3 Sample taking preservation and storage Pour the water sample into clean glass or plastic containers and analyse as soon as possible after the sample is taken If this is not possible fill th...

Page 134: ...AZINE test powder Note 3 to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Press key Wait for a reaction period of 10 minutes After reaction period is finished proceed as follows 8 Any slight turbidity that occurs when the reagent is added must be removed by filtration Note 4 9 Place the vial in the sample chamber making sure that the marks are...

Page 135: ...alitative folded filter papers for medium precipitates are recommend 5 In order to check whether the reagent has aged if it has been stored for a lengthy period perform the test as described above using tap water If the result is above the detection limit of 0 05 mg l you should only use the reagent with reservations major result deviation 6 There is an option to change the unit from mg l to μg l ...

Page 136: ...hat the marks are aligned 5 Press ZERO key 6 Remove the vial from the sample chamber 7 Fill the second clean vial with 10 ml of water sample this is the sample 8 Add 1 ml VARIO Hydra 2 Rgt Solution into the vial 9 Close the vial tightly with the cap and swirl several times to mix the contents 10 Place the vial the blank in the sample chamber making sure that the marks are aligned 11 Press TEST key...

Page 137: ...ease up to 20 Morpholine does not interfere up to 10 mg l Highly colored or turbid samples Mix 1 part deionized water with 1 part household bleach Add 1 drop of these mixture into 25 ml of water sample and mix Use 10 ml prepared sample in place of deionized water in point 1 Note at point 7 use the unprepared water sample Principle Hydrazine is oxidised by the household bleach The interference by c...

Page 138: ...maining in the beaker 6 Add one HYDROGENPEROXIDE LR tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablet 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 2 minutes After the reaction perio...

Page 139: ... must be avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for pH adjustment Strong alkaline or acidic water samples must be adjusted to between pH 6 and pH 7 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding t...

Page 140: ...GENPEROXIDE LR tablet straight from the foil and crush the tablet using a clean stir ring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the ...

Page 141: ... avoided The analysis must take place immediately after taking the sample 3 The DPD colour development is carried out at a pH value of 6 3 to 6 5 The reagent tablet therefore contains a buffer for the pH adjustment Strong alkaline or acid water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Exceeding of the meas...

Page 142: ...mber empty the vial leaving a view drops in 5 Add one DPD No 1 tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key The result is s...

Page 143: ...141 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Oxidising reagents such as Chlorine Bromine etc interfere as they react like Iodine ...

Page 144: ...f total dissolved Iron Fe2 and Fe3 Iron with Powder Pack 0 1 3 mg l Fe Determination of all soluble iron and most insoluble forms of iron Iron total with Powder Pack 0 1 1 8 mg l Fe Determination of all soluble iron and most insoluble forms of iron most insoluble iron oxides are recovered by the reagent This information refer to analysis of the water sample without digestion Further information yo...

Page 145: ...l 10 ml of this pre treated solution is used for the following analysis Perform as described at the selected test method 2 Water which has been treated with organic compounds like corrosion inhibitors must be oxidised where necessary to break down the iron Therefore add 1 ml concentrated sulfuric acid and 1 ml concentrated nitric acid to 100 ml of water sample and boil to approx half volume After ...

Page 146: ...f water sample 6 Add one IRON LR tablet straight from the foil and crush the tablet using a clean stirring rod Dissolve the tablet 7 Fill the 10 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically T...

Page 147: ...s the total dissolved Iron as Fe2 and Fe3 2 For the determination of Fe2 ions the IRON II LR tablet is used as described above instead of the IRON LR tablet 3 For the determination of total dissolved and undissolved iron digestion is required An example is described on page 143 ...

Page 148: ...of water sample 6 Add one IRON LR tablet straight from the foil and crush the tablet using a clean stirring rod Dissolve the tablet 7 Fill the 50 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically ...

Page 149: ...s the total dissolved Iron as Fe2 and Fe3 2 For the determination of Fe2 ions the IRON II LR tablet is used as described above instead of the IRON LR tablet 3 For the determination of total dissolved and undissolved iron digestion is required An example is described on page 143 ...

Page 150: ...ablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically T...

Page 151: ...s the total dissolved Iron as Fe2 and Fe3 2 For the determination of Fe2 ions the IRON II LR tablet is used as described above instead of the IRON LR tablet 3 For the determination of total dissolved and undissolved iron digestion is required An example is described on page 143 ...

Page 152: ... one Vario Ferro F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents Note 4 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 3 minutes Note 5 After the reaction period is finished the measurement starts automatically The result is show...

Page 153: ...ide requires a prior digestion use mild vigorous or Digesdahl digestion e g for digestion with acid see page 143 3 Very strong alkaline or acidic water samples must be adjusted to a ph Value between 3 and 5 before analysis 4 Accuracy is not affected by undissolved powder 5 Water samples containing visible rust should be allowed to react at least five minutes ...

Page 154: ...h vial 4 Close the vials tightly and swirl several times to mix the contents 5 Press key Wait for a reaction period of 3 minutes After reaction period is finished proceed as follows 6 Place the vial the blank in the sample chamber making sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Place the vial the sample in the sample chamber ma king sure that the...

Page 155: ...between pH 3 and pH 8 before the reagent is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Interferences When interferences occurred the colour development was inhibited or a precipitate was formed The values below refer to a standard with an iron concentration of 0 5 mg l The following substances do not interfere when present up to the levels given Substance no inerference to C...

Page 156: ...the cell and dry completely Caution Reagent Pb 1 contains Potassium Cyanide Adhere to the specified dosage sequence Note 4 5 Pipette 0 5 ml Reagent Pb 1 into a beaker 6 Add 0 5 ml Reagent Pb 2 and mix 7 Add 8 ml of water sample and mix 8 Fill the 10 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key The result...

Page 157: ... 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Volumes for sample and reagent should always metered by using volumetric pipettes class A 6 The test determines only Pb2 ions Samples must be pre treated or decomposed by digestion before colloidal undissolved and complex bounded l...

Page 158: ... l Pb Procedure A Select this test for the determination of lead in soft to medium hard waters with a Ca2 contents lower than 70 mg l approx 10 d Procedure B Select this test for the determination of lead in hard to very hard waters with a Ca2 contents between 70 mg l and 500 mg l approx 10 d to 70 d 2 3 4 2 3 5 ...

Page 159: ...pany MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature the sample temperature must be between 10 and 40 C 6 Sample volume should always metered by using volumetric pipette class A 7 The test determines only Pb2 ions Samples must be pre treated or decomposed by digestion before colloidal undissolved...

Page 160: ...4 4 Fill one reaction tube with the same size drops by hol ding the bottle vertically and squeeze slowly 5 drops reagent Pb 1K 5 Close the vial tightly with the cap and invert several times to mix the contents 6 Add 5 ml of water sample 7 Close the vial tightly with the cap and invert the vial several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are a...

Page 161: ...ightly with the cap and invert several times to mix the contents 6 Add 5 ml of water sample 7 Close the vial tightly with the cap and invert several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key 10 Remove the vial from the sample chamber and open carefully 11 Add 1 level microspoon of reagent Pb 2K 12 Close the vial tightly...

Page 162: ... clean stirring rod and dissolve the tablet 6 Add one MANGANESE LR 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 5 minutes...

Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...

Page 164: ...nt solution 6 Close the vials tightly and swirl several times to mix the contents 7 Fill each vial with the same size drops by holding the bottle vertically and squeeze slowly 21 drops of PAN Indicator solution 8 Close the vials tightly and swirl several times to mix the contents 9 Press key Wait for a reaction period of 2 minutes Note 4 After reaction period is finished proceed as follows 10 Plac...

Page 165: ...After adding the Vario Ascorbic Acid powder pack add additionally 10 drops of Rochelle Salt Solution 3 After addition of the reagent solution Alkaline Cyanide a cloudy or turbid solution may form in some water samples The turbidity should disappear after point 7 4 Water samples containing more than 5 mg l iron should be allowed to react at least 10 minutes 5 Conversion mg l MnO4 mg l Mn x 2 17 6 M...

Page 166: ...e the vial tightly with the cap and swirl several times to mix the contents 7 Add one VARIO Sodium periodate powder pack straight from the foil to the same water sample 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Place the vial in the sample chamber making sure that the marks are aligned 10 Press TEST key Wait for a reaction period of 2 minutes After the rea...

Page 167: ...ering capacity of the reagents and requires sample pre treatment If samples were acidified for storing adjust the pH between 4 and 5 with 5 mol l 5 N Sodium hydroxide before test Do not exceed pH 5 as manganese may precipitate 3 Interferences Interfering substance Interference level Calcium greater than 700 mg l Chloride greater than 70 000 mg l Iron greater than 5 mg l Magnesium greater than 100 ...

Page 168: ...tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 7 Add one MOLYBDATE HR No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 8 Dissolve the tablets using a clean stirring rod 9 Rinse out the vial with the prepared water sample and then fill to the 10 ml mark 10 Close the vial tightly with the ca...

Page 169: ...lets must be added in the correct sequence 2 Under test conditions pH 3 8 3 9 iron does not interfere nor do other metals at levels likely to be found in industrial water systems 3 Conversions mg l Mo mg l MoO4 x 0 6 mg l Na2 MoO6 mg l MoO4 x 1 3 4 MoO4 Mo Na2 MoO4 ...

Page 170: ...7 Add one Vario Molybdenum HR 2 F10 powder pack straight from the foil to the same water sample 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Add one Vario Molybdenum HR 3 F10 powder pack straight from the foil to the same water sample 10 Close the vial tightly with the cap and swirl several times to mix the contents 11 Place the vial in the sample chamber mak...

Page 171: ... nearly 7 with 1 mol l Nitric acid or 1 mol l Sodium hydroxide 3 Concentration from 10 mg l Cu causes too high test values if the described reaction time of 5 minutes is increased So it is very important to perform the test procedure continuously 4 Substances who may interfere when present in concentrations at Aluminium 50 mg l Chromium 1000 mg l Iron 50 mg l Nickel 50 mg l Nitrite all levels 5 Mo...

Page 172: ...water sample 6 Add 2 level spoons of No 8 black Nickel 51 and dissolve the reagent 7 Add 0 2 ml Nickel 52 to the same water sample and mix 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automa...

Page 173: ...171 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 The test sample and the reagents should have room temperature for test performance 2 The test sample should have a pH of between 3 and 9 ...

Page 174: ...to the prepared water sample 6 Close the vial tightly with the cap and invert several times to mix the contents 7 Add 0 2 ml Nickel 52 to the same water sample 8 Close the vial tightly with the cap and invert several times to mix the contents 9 Place the vial in the sample chamber making sure that the marks are aligned 10 Press TEST key Wait for a reaction period of 3 minutes After the reaction pe...

Page 175: ...173 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 The test sample and the reagents should have room temperature for test performance 2 The test sample should have a pH of between 3 and 9 ...

Page 176: ...nd invert the vials gently several times 10 x to mix the contents Note 1 5 Press key Wait for a reaction period of 5 minutes 6 After reaction period is finished proceed as follows 7 Place the vial the blank in the sample chamber making sure that the marks are aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Place the vial the sample in the sample chamber making sure that the m...

Page 177: ...175 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Some solids may not dissolve 2 Conversion mg l NO3 mg l N x 4 43 3 N NO3 ...

Page 178: ...p and invert several times to mix the contents Caution tube becomes warm 6 Add 0 2 ml Nitrate 111 7 Close the vial tightly with the cap and invert several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 15 minutes After the reaction period is finished the measurement starts automatically The resu...

Page 179: ...177 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Nitrite concentrations greater than 2 mg L NO2 lead to higher test results 2 Great quantities of COD lead to higher test results 3 N NO3 ...

Page 180: ...let straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 10 minutes After the reaction period is finished the measurement starts automatically Th...

Page 181: ...mony III Iron III Lead Mercury I Silver Chloroplatinate Metavanadate and Bismuth Copper II ions may cause lower test results as they accelerate the decomposition of the Diazonium salt It is improbable in practice that theses interfering ions will occur in such high concentrations that they cause significant reading errors 2 Conversion mg l NO2 mg l N x 3 29 3 N NO2 ...

Page 182: ... 5 Add one VARIO Nitri 3 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 20 minutes After the reaction period is finished the measurement starts automatically The result is shown in the ...

Page 183: ...us Auric Bismuth Chloroplatinate Ferric Lead Mercurous Metavanadate Silver ions interfere by causing precipitation In samples with very high concentrations of Nitrate 100 mg l N a small amount of Nitrite will be found Such high levels of Nitrate appear to undergo a slight amount of reduction to Nitrite either spontaneously or during the reaction time of the test 2 N NO2 ...

Page 184: ...p and invert several times to mix the contents 6 Add one level spoon of No 8 black Nitrite 101 7 Close the vial tightly with the cap and swirl until the reagent is solved completely 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reaction period is finished the measurement starts automatically The res...

Page 185: ... Spectrophotometer 01 2008 1 1 Methods Notes 1 Store the reagents in closed containers at a temperature of 4 C to 8 C 2 The test sample and the reagents should have room temperature for test performance 3 N NO2 ...

Page 186: ...p and invert several times to mix the contents 6 Add one level spoon of No 8 black Nitrite 101 7 Close the vial tightly with the cap and swirl until the reagent is solved completely 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reaction period is finished the measurement starts automatically The res...

Page 187: ... Spectrophotometer 01 2008 1 1 Methods Notes 1 Store the reagents in closed containers at a temperature of 4 C to 8 C 2 The test sample and the reagents should have room temperature for test performance 3 N NO2 ...

Page 188: ...owder Pack to each vial Note 2 8 Close the vials with the caps and shake to mix the contents at least 15 seconds 9 Press key Wait for a reaction period of 3 minutes After reaction period is finished proceed as follows 10 Open the digestion vials and add one Vario TN Reagent B powder pack to each vial Note 2 11 Close the vials with the caps and shake to mix the contents at least 15 seconds Note 8 1...

Page 189: ...r exactly 30 minutes 8 The reagent will not completely dissolve 9 Hold the vial in a vertical position with the cap pointing up Turn the vial upside down Wait for all of the solution to flow down to the cap Return the vial to the upright position Wait for all the solution to flow to the bottom of the vial This process is one inversion 10 inversions approx 30 seconds 10 After zero calibration with ...

Page 190: ...Powder Pack to each vial Note 2 8 Close the vials with the caps and shake to mix the contents at least 15 seconds 9 Press key Wait for a reaction period of 3 minutes After reaction period is finished proceed as follows 10 Open the digestion vials and add one Vario TN Reagent B powder pack to each vial Note 2 11 Close the vials with the caps and shake to mix the contents at least 15 seconds Note 8 ...

Page 191: ...actly 30 minutes 8 The reagent will not completely dissolve 9 Hold the vial in a vertical position with the cap pointing up Turn the vial upside down Wait for all of the solution to flow down to the cap Return the vial to the upright position Wait for all the solution to flow to the bottom of the vial This process is one inversion 10 inversions approx 30 seconds 10 After zero calibration with the ...

Page 192: ...e hot Invert the vial and allow to cool to room temperature 6 Add 1 level spoon of No 4 grey compensation reagent 7 Close the vial tightly with the cap and invert several times to mix the contents 8 Use this pre treated sample for the following test procedure Performing test procedure 9 Place the supplied blank red label in the sample cham ber making sure that the marks are aligned 10 Press ZERO k...

Page 193: ... 15 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l total Nitrogen Notes 1 This test determines the inorganic compounds Ammonium Nitrate and Nitrite as well as organic compounds like amino acid urea complexing agents etc 2 Nitrogen compounds which are hardly to oxidise as may be found in industrial sewage are not digeste...

Page 194: ...N The vials are hot Invert the vial and allow to cool to room temperature 6 Add 1 level spoon of No 4 grey compensation reagent 7 Close the vial tightly with the cap and invert several times to mix the contents 8 Use this pre treated sample for the following test procedure Performing test procedure 9 Place the supplied blank red label in the sample cham ber making sure that the marks are aligned 1...

Page 195: ... 15 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l total Nitrogen Notes 1 This test determines the inorganic compounds Ammonium Nitrate and Nitrite as well as organic compounds like amino acid urea complexing agents etc 2 Nitrogen compounds which are hardly to oxidise as may be found in industrial sewage are not digeste...

Page 196: ...let straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The...

Page 197: ...ive Oxygen is a synonym for a common disinfectant based on Oxygen in Swimming Pool Treatment 1 When preparing the sample the escape of Oxygen gases e g by pipetting or shaking must be avoided 2 The analysis must take place immediately after taking the sample ...

Page 198: ... 0 02 1 mg l O3 The following selection is shown in the display for the determination of Ozone in the presence of Chlorine for the determination of Ozone in the absence of Chlorine Select the desired method with the arrow keys and Confirm with key Ozon with Cl without Cl with Cl without Cl 2 9 9 3 0 0 ...

Page 199: ...er for the pH adjustment Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before the tablet is added use 0 5 mol l Sulfuric acid resp 1 mol l Sodium hydroxide 4 Turbidity lead to errors The use of the DPD No 1 tablet in samples with high Calcium ion contents and or high conductivity can lead to turbidity of the sample and therefore incorrect measurements it is not pos...

Page 200: ...e tablets 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically 11 Remove the cell from the sample chamber Empty the cell and dry completely 12 Rinse a beaker with the sample and empty ...

Page 201: ...solution 17 Place the cell in the sample chamber making sure that the positioning is correct 18 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l Ozone mg l total Chlorine Notes see page 197 mg l O3 mg l tot Cl T1 accepted prepare T2 press TEST Countdown 2 00 ...

Page 202: ... in the beaker 6 Add one DPD No 1 tablet and one DPD No 3 tablet straight from the foil and crush the tablet using a clean stirring rod 7 Add 10 ml of water sample and dissolve the tablets 8 Fill the 50 mm cell with the colored test solution 9 Place the cell in the sample chamber making sure that the positioning is correct 10 Press TEST key Wait for a reaction period of 2 minutes After the reactio...

Page 203: ...201 SP600 Spectrophotometer 01 2008 ...

Page 204: ... 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically 10 Remove the vial from the sample chamber Empty the vial ...

Page 205: ...tents of the second vial into the prepared vial 16 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 17 Place the vial in the sample chamber making sure that the marks are aligned 18 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l Oz...

Page 206: ... one DPD No 3 tablet straight from the foil and crush the tablets using a clean stirring rod 6 Add water sample to the 10 ml mark 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the...

Page 207: ...205 SP600 Spectrophotometer 01 2008 ...

Page 208: ... vial from the sample chamber 5 Add one PHENOL RED PHOTOMETER tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value prepar...

Page 209: ... can produce results inside the measuring range A plausibility test pH meter is recommended 4 The accuracy of the colorimetric determination of pH values depends on various boundary conditions buffer capacity of the sample salt contents etc 5 Salt error Correction of test results average values for samples with salt contents of Indicator Salt contents Phenol red 1 molar 0 21 2 molar 0 26 3 molar 0...

Page 210: ... key 4 Remove the vial from the sample chamber 5 Fill the vial with the same size drops by holding the bottle vertically and squeeze slowly 6 drops of PHENOL RED solution 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display as pH value prepa...

Page 211: ...rystal of Sodiumthiosulfate Na2 S2 O3 x 5 H2 O to the sample before adding the PHENOL RED solution PHENOL RED tablets already contain Thiosulfate 2 Due to differing drop size results can show a discrepancy in accuracy by comparison with tablets This can be minimised by using a pipette 0 18 ml PHENOL RED solution is equivalent to 6 drops 3 After use replace the bottle cap securely 4 Store the reage...

Page 212: ...let using a clean stirring rod 6 Add one PHENOLE No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl the vial several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 5 minutes After...

Page 213: ...rent types of phenolic compounds the result is displayed as the equivalent concentration of Phenol C6 H5 OH 2 The test sample should have a pH of between 3 and 11 3 Interferences can be caused in the presence of reducing agents oxidising reagents sulphides or suspended solids Distillation of the sample is necessary then see Standard Methods for Examination of Water and Wastewater 20th Edition 4 40...

Page 214: ...s Phosphate ortho LR with Tablet 0 05 4 mg l PO4 Determination of ortho Phosphate ions Phosphate ortho HR with Tablet 1 80 mg l PO4 Determination of ortho Phosphate ions Phosphate ortho with Powder Pack 0 06 2 5 mg l PO4 Determination of ortho Phosphate ions Phosphate ortho with Tube Test 0 06 5 mg l PO4 Determination of ortho Phosphate ions Phosphate ortho Vanadat Molybdat with Tube Test 3 60 mg ...

Page 215: ...thod 321 and 322 the ortho Phosphate ions react with the Vanadate molybdate reagent under acid conditions to a yellow colored product Notes only for tube tests and tests with Powder Packs 323 324 325 326 1 Application for water wastewater and seawater 2 Highly buffered samples or samples with extreme pH Values should be adjusted between pH 2 and pH 10 before analysis with 1 mol l Hydrochloric acid...

Page 216: ...um hydroxide solution to the vial 7 Close the vial with the cap and invert the vial gently several times to mix the contents 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press ZERO key 10 Remove the vial from the sample chamber 11 Add one VARIO Phosphate Rgt F10 powder pack straight from the foil to the vial Note 2 12 Close the vial tightly with the cap and swirl...

Page 217: ...propriate safety precautions and a good lab technique should be used during the whole procedure 2 Use a funnel to add the reagent 3 The reagent dissolves not completely 4 See also page 213 5 Conversions mg l PO4 mg l P x 3 07 mg l P2 O5 mg l P x 2 29 6 P PO4 P2 O5 ...

Page 218: ...label in the sample cham ber making sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Add 2 drops 0 1 ml Phosphate 101 into the prepa red sample see step 5 10 Close the vial tightly with the cap and invert several times to mix the contents 11 Add one level spoon of No 4 grey Phosphate 102 12 Close the vial tightly with the cap and swirl until the reagent ...

Page 219: ...217 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 If the analysis is performed without digestion only ortho Phosphate ions are determined 2 See also page 213 3 P PO4 P2 O5 ...

Page 220: ... label in the sample cham ber making sure that the marks are aligned 7 Press ZERO key 8 Remove the vial from the sample chamber 9 Add 2 drops 0 1 ml Phosphate 101 into the prepa red sample see step 5 10 Close the vial tightly with the cap and invert several times to mix the contents 11 Add one level spoon of No 4 grey Phosphate 102 12 Close the vial tightly with the cap and swirl until the reagent...

Page 221: ...219 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 If the analysis is performed without digestion only ortho Phosphate ions are determined 2 See also page 213 3 P PO4 P2 O5 ...

Page 222: ...blet using a clean stirring rod 6 Add one PHOSPHATE No 2 LR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the...

Page 223: ...ect sequence 3 The test sample should have a pH Value between 6 and 7 4 Interferences Higher concentrations of Cu Ni Cr III V V and W VI interfere due to their colour Silicates doe not interfere masked by Citric acid in the tablets 5 See also page 213 6 Conversion mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 7 PO4 P P2 O5 ...

Page 224: ...e tablet using a clean stirring rod 6 Add one PHOSPHATE HR P2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After...

Page 225: ...or samples under 5 mg l PO4 it is reccommendable to analyse the water sample with e g method 320 Posphate LR ortho with Tablet 2 Only ortho Phosphate ions react 3 See also page 213 4 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 5 PO4 P P2 O5 ...

Page 226: ...Phosphate Rgt F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 10 15 sec Note 1 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result ...

Page 227: ...225 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 The reagent dissolves not completely 2 See also page 213 3 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 4 PO4 P P2 O5 ...

Page 228: ...0 powder pack straight from the foil to the water sample Note 1 6 Close the vial tightly with the cap and swirl several times to mix the contents approx 10 15 sec Note 2 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 2 minutes After the reaction period is finished the measurement starts automatically The result is shown ...

Page 229: ...pectrophotometer 01 2008 1 1 Methods Notes 1 Use a funnel to add the reagent 2 The reagent dissolves not completely 3 See also page 213 4 Conversions mg l P mg l PO4 x 0 33 mg l P2 O5 mg l PO4 x 0 75 5 PO4 P P2 O5 ...

Page 230: ...ter sample into one reaction tube 5 Close the vial tightly with the cap and invert several times to mix the contents 6 Place the vial in the sample chamber making sure that the marks are aligned 7 Press TEST key Wait for a reaction period of 3 minutes After the reaction period is finished the measurement starts automatically The result is shown in the display in mg l ortho Phos phate Ø 16 mm prepa...

Page 231: ...229 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 Only ortho Phosphate ions react 2 P PO4 P2 O5 ...

Page 232: ...e vial with the cap and invert the vial gently several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press ZERO key 9 Remove the vial from the sample chamber 10 Add one VARIO Phosphate Rgt F10 powder pack straight from the foil to the vial Note 2 11 Close the vial tightly with the cap and swirl several times to mix the contents approx 10 ...

Page 233: ...propriate safety precautions and a good lab technique should be used during the whole procedure 2 Use a funnel to add the reagent 3 The reagent dissolves not completely 4 See also page 213 5 Conversions mg l PO4 mg l P x 3 07 mg l P2 O5 mg l P x 2 29 6 PO4 P P2 O5 ...

Page 234: ...Remove the vial from the sample chamber 5 Add one Potassium T tablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key The result is shown in the display in mg l Potassium ...

Page 235: ...233 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 If Potassium is present a cloudy solution will be given Single particles are uncaused necessarily by Potassium ...

Page 236: ...ore width of 0 45 μm At least 100 ml of the water sample should be filtered Performing test procedure 2 Fill a clean 50 mm cell with deionized water Note 1 3 Place the cell in the sample chamber making sure that the positioning is correct 4 Press ZERO key 5 Remove the cell from the sample chamber and empty the cell 6 Use some of the filtered water sample to rinse out the cell then fill the sample ...

Page 237: ...variable used to describe the true coloration of a water sample The true coloration of a water sample is the coloration caused solely by dissolved substances in the sample This is why the water sample has to be filtered prior to measurement Measurement at a wavelength of 436 nm is obligatory and is adequate for natural waters and the outflow of municipal sewage plants As industrial waste waters of...

Page 238: ...iod of 5 minutes After reaction period is finished proceed as follows 8 Add one SILICA PR tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 9 Add one SILICA No 2 tablet straight from the foil to the same water sample and crush the tablet using a clean stirring rod 10 Close the cap tightly and swirl the vial several times until the tablets are di...

Page 239: ...The tablets must be added in the correct sequence 2 Phosphate ions does not interfere under the given reaction conditions 3 If Phosphate is known to be absent the addition of the SILICA PR tablet may be omitted 4 Conversion mg l Si mg l SiO2 x 0 47 5 SiO2 Si ...

Page 240: ...eaction period is finished proceed as follows 5 Add one Vario Silica Citric Acid F10 Powder Pack straight from the foil into each vial 6 Close the vials tightly and swirl several times to mix the contents 7 Press key Wait for a reaction period of 1 minute Note 3 After reaction period is finished proceed as follows 8 Place the vial the blank in the sample chamber making sure that the marks are alig...

Page 241: ... time of 8 minutes is required 3 The given reaction time of 1 minute refers to a water sample temperature of 20 C At 30 C a reaction time of 30 seconds at 10 C a reaction time of 2 minutes is required 4 Interferences Substance Interference Iron large amounts interfere Phosphate does not interfere at concentrations less than 50 mg l PO4 at 60 mg l PO4 the interference is approx 2 at 75 mg l PO4 the...

Page 242: ...l to the same water sample Note 2 8 Close the vial with the caps tightly and swirl the vials several times to mix the contents 9 Press key Wait for a reaction period of 10 minutes After reaction period is finished proceed as follows 10 Add one Vario Silica Citric Acid F10 powder pack straight from the foil to the water sample Note 3 11 Close the vial tightly with the cap and swirl several times to...

Page 243: ...ations less than 50 mg l PO4 at 60 mg l PO4 the interference is approx 2 at 75 mg l PO4 the interference is approx 11 Sulfide interfere at all levels Occasionally water samples contain silica forms which reacts very slowly with Molybdate The nature of these forms is not known A pre treatment with Sodium hydrogencarbonate and then with Sulfuric Acid will make these forms reactive to Molybdate pre t...

Page 244: ...Add one VARIO Sulpha 4 F10 powder pack straight from the foil to the water sample 6 Close the vial tightly with the cap and swirl several times to mix the contents 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically The result is shown in the...

Page 245: ...243 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 If Sulfate ions are present a cloudy solution will be given ...

Page 246: ... clean stirring rod and dis solve the tablet 6 Add one SULFIDE No 2 tablet to the same water sample and crush the tablet using a clean stirring rod 7 Close the vial tightly with the cap and swirl several times until the tablets are dissolved 8 Place the vial in the sample chamber making sure that the marks are aligned 9 Press TEST key Wait for a reaction period of 10 minutes After the reaction per...

Page 247: ...eact with DPD do not interfere in the test 3 To avoid loss of Sulfide collect the sample carefully with a minimum of aeration It is essential to test the sample immediately after collection 4 The temperature of test performance should be 20 C Difference to this temperature can lead to higher or lower results 5 Conversion H2 S mg l S x 1 06 6 S H2 S ...

Page 248: ... water sample 6 Add one SULFITE LR tablet straight from the foil and crush the tablet using a clean stirring rod Dissolve the tablet 7 Fill the 10 mm cell with the colored test solution 8 Place the cell in the sample chamber making sure that the positioning is correct 9 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically...

Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...

Page 250: ...ablet straight from the foil to the water sample and crush the tablet using a clean stirring rod 6 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 7 Place the vial in the sample chamber making sure that the marks are aligned 8 Press TEST key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically T...

Page 251: ...249 SP600 Spectrophotometer 01 2008 ...

Page 252: ... tube with the same size drops by holding the bottle vertically and squeeze slowly 3 drops reagent T 1K Do not mix contents 6 Add 3 drops reagent T 2K 7 Close the vial tightly with the cap and shake for 30 seconds 8 Press key Wait for a reaction period of 10 minutes After reaction period is finished proceed as follows Swirl the vial and than place the vial in the sample chamber making sure that th...

Page 253: ... 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Because reaction depends on temperature sample and tube temperature must be between 10 and 20 C 6 Sample volume should always metered by using volumetric pipette class A 7 MBAS Methylene Blue Active Substances calculated as sodium ...

Page 254: ...estion 6 Pipette 3 ml pre prepared sample into one reaction tube 7 Add 1 level microspoon of reagent TOC 2K 8 Immediately close the vial with an aluminium cap tightly 9 Heat vials standing on its head at 120 C in the pre heated reactor for 120 minutes 10 Wait for 1 hour before proceeding Do not cool down with water Performing test procedure 11 Place the supplied blank in the sample chamber making ...

Page 255: ...instructions delivered with the test and the MSDS available at www merck com 3 Spectroquant is a registered trade mark of the company MERCK KGaA 4 Appropriate safety precautions and good lab technique should be used during the whole procedure 5 Sample volume should always metered by using volumetric pipette class A 6 TOC Total Organic Carbon ...

Page 256: ...oning is correct 3 Press ZERO key 4 Remove the cell from the sample chamber and empty completely 5 Rinse the 50 mm cell with the water sample and then fill with water sample 6 Place the cell in the sample chamber making sure that the positioning is correct 7 Press TEST key The result is shown in the display in FAU prepare Zero press ZERO Zero accepted prepare Test press TEST 3 8 5 ...

Page 257: ...measurements The attenuated radiation method is different from the Nephelometric method 2 Colour interference is minimized through measurement at 860 nm Interferences depend on light absorption at 860 nm and gas bubbles 3 Collect water samples in clean glass bottles and analyse the water sample as soon as possible It is possible to store the sample at 4 C for 48 hours Before measurement warm up th...

Page 258: ...h the cap and swirl several times to mix the contents 7 Add 1 drop of Urea Reagent 2 Urease to the same water sample Note 8 8 Close the vial tightly with the cap and swirl several times to mix the contents 9 Press key Wait for a reaction period of 5 minutes After reaction period is finished proceed as follows 10 Add one AMMONIA No 1 tablet straight from the foil to the prepared water sample and mi...

Page 259: ...nation at the latest one hour after sample taking 3 The tablets must be added in the correct sequence 4 Store reagent 2 Urease in the refrigerator at a temperature of 4 C to 8 C 5 The AMMONIA No 1 tablet will only dissolve completely after the AMMONIA No 2 tablet has been added 6 Ammonium and chloramines are also measured during urea measurement 7 Before analysing seawater samples a measuring spoo...

Page 260: ...marks are aligned 5 Press ZERO key Wait for a reaction period of 5 minutes After the reaction period is finished the measurement starts automatically 6 Remove the vial from the sample chamber 7 Add one EDTA tablet straight from the foil to the prepared vial and crush the tablet using a clean stirring rod 8 Close the vial tightly with the cap and swirl several times until the tablet is dissolved 9 ...

Page 261: ...2 In the case of high levels of residual chlorine perform the analysis with a dechlorinated water sample To dechlorinate add one DECHLOR tablet to the water sample point 1 Crush and mix to dissolve the tablet Then add the COPPER ZINC LR tablet point 2 and continue with the test procedure as described above ...

Page 262: ...let reagents should be added to the water sample straight from the foil without touching them with the fingers Liquid reagents Add the same size drops to the water sample by holding the bottle vertically and squeezing slowly After use replace the bottle caps securely noting the colour coding Note recommendation for storage e g cool and dry Powder Packs 1 2 3 4 ...

Page 263: ...e Clean vials and accessories after each analysis as soon as possible a Clean vials and accessories with laboratory detergent e g Extran MA 02 neutral phosphatic Extran MA 03 alkaline phosphate free from Merck KGaA b Rinse with tap water thoroughly c On demand see Notes perform special cleaning at this point e g rinse with diluted Hydrochloric acid solution d Rinse with deionized water thoroughly ...

Page 264: ...ith the mark on the vial aligned with the mark on the instrument 5 Always perform zeroing and test with closed photometer lid 6 Bubbles on the inside wall of the vial leads to incorrect measurements To prevent this remove the bubbles by swirling the vial before performing the test 7 Avoid spillage of water in the sample chamber If water should leak into the instrument housing it can destroy electr...

Page 265: ...methods Caution 1 Dilution decreases accuracy 2 Do not dilute water samples for measurement of pH values This will lead to incorrect test results If there is displayed Overrange use another instrument e g pH meter 1 2 5 Correcting for volume additions If a larger volume of acid or base is used to pre adjust the pH value a volume correction of the displayed result is necessary Example For adjusting...

Page 266: ...264 SP600 Spectrophotometer 01 2008 ...

Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...

Page 268: ...s and photometer setting if there is no power from the power supply from the mains adapter As long as the instrument is energised from the mains adapter the batteries not discharged Recommendation Exchange of the batteries every 3 years When no mains adapter supplies energy to the instrument all stored data and settings will be lost if the batteries are taken out Recommendation Keep the instrument...

Page 269: ...hen insert the cell so that a matt side points towards the viewer 20 30 40 mm and other rectangular cells Insert the cell always in contact to the clip C on the left side of the cell chamber 50 mm cells no cell holder required Cylindric cells 16 and 24 mm diameter Note Cylindric cells are called vials in the test instructions Vials are placed as indicated in the cell holder the 2 marks matching ea...

Page 270: ...unction key description in the text if key available Confirming Menu of photometer settings and further functions Moving the cursor up resp down Storing of displayed test result Performing Zero Performing Test Displaying date and time user countdown 2 2 2 Displaying time and date Press clock key The display shows After 15 seconds the photometer reverts to the previous display automatically or pres...

Page 271: ...s key to accept the last used user count down or press any number key to start entering a new value The entering comprises two digits each Enter minutes and seconds e g 2 minutes 0 seconds 0 2 0 0 Confirm with key The display shows Start count down with key After countdown has finished the photometer reverts to the previous display automatically 19 30 20 2006 06 15 Countdown mm ss 99 99 Countdown ...

Page 272: ... shows get sure that the cell chamber is empty and the photometer lid closed Start selftest by pressing key now the photometer is performing the selftest for approx 21 2 minutes During this time the following is checked function of the tungsten halogen lamp function of the step motor accuracy of wavelength with internal Didymium filter and if necessary adjustment in this case the selftest can be t...

Page 273: ...ponding ranges Changing chemical species see chapter 2 3 7 page 274 Line 1 Method number Method name Line 2 Range with chemical species 1 Line 3 Range with chemical species 2 Line 4 Range with chemical species 3 320 Phosphate LR T 0 05 4 mg l PO4 0 02 1 3 mg l P 0 04 3 mg l P2 O5 Method Method Information Methodlist Chemical Species Information Information F1 F2 F1 F1 F2 F2 2 3 2 Selecting a metho...

Page 274: ...tion Press arrow key or to select the required determi nation Confirm with key 2 3 4 Performing Zero The display shows Prepare a clean vial as described in Method and place the vial in the sample chamber making sure that the marks are aligned Press ZERO key The display shows prepare Zero press ZERO Zero accepted prepare Test press TEST Chlorine diff free total Zero ...

Page 275: ...untdown with key and proceed as described in the mode description The vial must not be placed in the sample chamber Press TEST key Prepare the water sample as described in the method description and place the vial in the sample chamber The display shows the countdown by pressing the TEST key and the countdown is started automatically The reaction period is finished the measurement starts automatic...

Page 276: ...nd will be displayed if this method is used the next time If there is the possibility to change the chemical species for a method it is described in the manual The arrows with the possible chemical species are printed below the notes of the method PO4 P P2 O5 2 3 8 Storing results Press STORE key during the test result is displayed The display shows We advise you to enter a numeric code up to 6 pl...

Page 277: ...results If a printer is installed and switched on it is possible to print out the test results without saving it before Press F3 key The entire data set is printed with date time Code No method and test result Printing example 100 Chlorine T 0 02 6 mg l Cl2 Profi Mode no 2006 07 01 14 53 09 Test No 1 Code Nr 007 4 80 mg l Cl2 The test No is an internal number that is set automatically if a test re...

Page 278: ...ay shows Confirm with TEST key or Press ZERO key to perform a new zero calibration The display shows 2 3 11 Selecting a new method Press ESC key to return to method selection Or enter the required method number directly e g 1 6 0 for Cyanuric acid Confirm with key Zero accepted prepare Test press TEST prepare Zero press ZERO Test Zero Esc 1 6 0 Test ...

Page 279: ...contrast 80 Setting the display contrast 316 LCD brightness 81 Setting the display brightness 316 Method list 60 User method list adaption 304 M list all on 61 User method list switching on all methods 305 M list all off 62 User method list switching off all methods 305 Print 20 Printing all stored results 282 Print code Nr 22 Print only results of a selected Code No range 284 Print date 21 Print ...

Page 280: ... Temperature 71 Selection of C or F for Langelier Mode 70 315 User calibration 45 Storage user calibration 294 User concentration 64 Entering of the data that are necessary to run a user concentration method 306 User polynoms 65 Entering of the data that are necessary to run a user polynomial 308 User methods clear 66 Delete all data of a user polynomial or of a concentration method 311 User metho...

Page 281: ...ountdown even if the key beep is switched off Setting date and time Press MODE 1 2 keys Confirm with key The display shows The entering comprises two digits each Enter year month and day e g 14 May 2006 0 6 0 5 1 4 Enter hours and minutes e g 3 07 p m 1 5 0 7 Confirm with key Note While conforming date and time with key the seconds are adjusted to zero automatically Key Beep ON 1 OFF 0 clock yy mm...

Page 282: ...y shows Press 0 key to switch the countdown off Press 1 key to switch the countdown on Confirm with key Notes 1 It is possible to finish the working countdown by pressing the key application e g serial analysis The user countdown is also available if the countdown is switched off 2 If the countdown function is switched off the operator is responsible for ensuring the necessary reaction period by h...

Page 283: ...hort beep Press MODE 1 4 keys Confirm with key The display shows Press 0 key to switch the signal beep off Press 1 key to switch the signal beep on Confirm with key Note In the case of methods with reaction periods an acoustic signal still sounds during the last 10 seconds of the countdown even if the key beep is switched off Signal Beep ON 1 OFF 0 Mode 1 4 0 1 ...

Page 284: ...onfirm with key The display shows Press key for printing out all stored test results The display shows e g After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entering by ESC All stored data are printed out Printing Print all Data Start cancel ESC Test No Mode 2 0 ...

Page 285: ...ast day of the required period e g 19 May 2006 0 6 0 5 1 9 Confirm with key The display shows Press key and all stored results in the selected date range are printed After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entering by ESC If you want to print only results of one day enter the same date twice to characterise the period Print sorted date f...

Page 286: ...Code No e g 1 0 Confirm with key The display shows Press key and all stored results in the selected Code Number range are printed After printing the photometer goes back to mode menu automatically Note It is possible to cancel the entering by ESC If you want to print only results of one Code Number enter the same Code Number twice If you want to print all results without Code No Code Nr is 0 enter...

Page 287: ...od number directly Confirm with key In case of differentiated methods select the required kind of determination and confirm with key The display shows Press key and all stored results of the selected method are printed After printing the photometer goes back to mode menu automatically Print 20 Acid demand 30 Alkalinity total 40 Aluminium T Print method 30 Alkalinity total Start cancel ESC Mode 2 3...

Page 288: ...key or to select the required Protocol Xon Xoff Hardware no control Confirm with key Finish with ESC key Flow Control will be set to the selection displayed at is Press 2 key to select Baud rate The display shows printing parameter 1 Flow control 2 Baud rate cancel ESC Flow Control is Hardware select save cancel ESC Baud rate is 19200 select save cancel ESC Mode 2 9 2 1 Esc ...

Page 289: ...ss arrow key to display the following test result Press arrow key to display the previous test result If there are no test results in memory the display shows Press arrow key or to select the required baud rate 600 1200 2400 4800 9600 14400 19200 Confirm with key End with ESC key Back to mode menu with ESC key Back to method selection with ESC key Note Select Hardware as Flow control and 9600 as b...

Page 290: ...d e g 19 May 2006 0 6 0 5 1 9 Confirm with key The display shows Press key and all stored results in the selected date range are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Note It is possible to cancel the entering by ESC If you want to recall only results of one day enter the same date twice to characterise the time period Storage ...

Page 291: ...key The display shows Press key and all stored results in the selected Code No range are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Note It is possible to cancel the entering by ESC If you want to recall only results of one Code Number enter the same Code Number twice If you want to recall all results without Code No Code Nr is 0 en...

Page 292: ... key In case of differentiated methods select the required kind of determination and confirm with key The display shows Press key and all stored results of the selected method are displayed Press F3 key to print the displayed result Press F2 key to print all selected results End with ESC Storage 20 Acid demand 30 Alkalinity total 40 Aluminium T Storage method 30 Alkalinity total Start cancel ESC p...

Page 293: ...eleting data by pressing ESC key Note All stored test results are deleted 2 4 5 Calibration Calibration Fluoride Regard notes Press MODE 4 0 keys Confirm with key The display shows 1 Fill a clean vial 24 mm Ø with exact 10 ml of deionized water close tightly with the cap 2 Place the vial in the sample chamber making sure that the marks are aligned Delete data Delete all data YES 1 NO 0 Calibration...

Page 294: ...lled up to the top 11 Place the vial in the sample chamber making sure that the marks are aligned 12 Press TEST key The display shows Confirm with key Back to method selection with ESC key Select method Fluoride with keys 1 7 0 und Note The same batch of SPADNS reagent solution must be used for adjustment and test The adjustment process needs to be performed for each new batch of SPANDS reagent so...

Page 295: ...ge the displayed results to the required value If a method use a mathematic equation for the calculation of the result it is only possible to calibrate the basic tests since all the other tests use the same polynom The same applies for some test procedures which use a polynom of another test procedure Please find information about useful calibration points at our homepage in the download area Retu...

Page 296: ...row key once decreases the displayed result Press keys till the displayed result corresponds to the value of the standard Confirm with key to store the new calibration factor Cancel user calibration by pressing ESC key The display shows Now the method name is displayed inverse and the test result is calculated with the new calibration factor 100 Chlorine T 0 02 6 mg l Cl2 0 90 mg l free Cl2 Jus Fa...

Page 297: ...roing the instrument press MODE 4 6 keys and confirm with key The display shows Press 1 key to delete user calibration Press 0 key to keep the valid user calibration The instrument goes back to Zero query automatically user calibration 100 Chlorine T 0 02 6 mg l Cl2 clear user calibration YES 1 NO 0 100 Chlorine T 0 02 6 mg l Cl2 prepare ZERO press ZERO 1 0 Mode 4 6 ...

Page 298: ...rovides only a minimum of operator instructions The criteria specified above d e f are not longer included Press MODE 5 0 keys in succession Confirm with key The display shows Press 0 key to switch the Profi Mode off Press 1 key to switch the Profi Mode on The display shows Confirm with key Note Storage of test results is possible In case of stored test results the display shows Profi Mode additio...

Page 299: ...into the sample chamber e g cell with deionized water or reagent blank Press ZERO key Place the cell with the test solution into the sample chamber Press TEST key The result is shown in the display as Extinction in Abs and Transmission in Abs Trans wavelength _ _ _ nm wavelength 540 nm prepare Zero press ZERO wavelength set Zero accepted prepare Test press TEST Abs Trans wavelength 540 nm E 0 596 ...

Page 300: ... 4 0 0 and confirm with key Enter end wavelength e g 6 2 1 and confirm with key The display shows Place a filled cell the blank into the sample chamber Note1 Press ZERO to start null balance the baseline The display shows Place the cell with the test solution into the sample cham ber Press TEST key Spectrum Start _ _ _ nm Spectrum End _ _ _ nm Spectrum 400 621 nm prepare Zero press ZERO Zero accep...

Page 301: ...ll print the result list nm mAbs or transfer the data to a PC using HyperTer minal When in peak valley list the F3 key will print the peak valley data Press ESC key to get back to wavelength selection Press ESC again to get back to mode menu Notes 1 It is possible to perform the baseline against air For the measurement of water samples it is recommend to use a cell filled with deionized water 2 10...

Page 302: ...0 2 0 Kinetic With method 54 Kinetic it is possible to describe the time dependency of a reaction e g time of colour development The maximum number of measuring points is 199 with an interval time of 6 to 999 seconds Concentrations of unknown samples can be calculated by a known factor The factor calculation is placed right before the sample measurement To evaluate the factor the slope of a best f...

Page 303: ...er pressing 1 key a factor can be entered After pressing 2 key a standard of known concentra tion can be measured Option Factor Enter a known factor with max 6 digits after decimal point in scientific notation Note 2 Press or key to change between plus and minus sign Enter the factor value with decimal point e g 9 3 4 Confirm with key Enter the factors exponent Press or key to change between plus ...

Page 304: ...t key Flow of Standard Sample Measurement Note 4 The display shows Place a zero vial in the sample chamber Press Zero key The display shows Place a prepared sample vial into the sample chamber Press Test key Note If a standard measurement has been performed before the sample measurement the following lines are displayed If the same zero shall be used press Test key To measure a new Zero press Zero...

Page 305: ...rbances in Abs at the time points T after x seconds To scroll through the values use the arrow keys or The value table can be printed or transmitted to a PC use of Hyperterminal by pressing F3 key Press Esc key to initialise a new measurement To start a new sample measurement press Test key Notes By pressing F1 key the displayed measurement result can be switched from 1 Absorbance to Transmission ...

Page 306: ...isplay shows Start with key The complete method list is displayed Methods with a point behind the method number will be displayed in the method selection list Methods without a point will not be displayed in the method selection list Press key or to select the required method from the displayed list Switch with F2 key between active and inactive Select next method activate or inactivate it and so ...

Page 307: ... instrument goes back to mode menu automatically User method list switch all methods off The program structure requires that the method list must have at least one active switched on method For this reason the instrument activates one method automatically Press MODE 6 2 keys Confirm with key The display shows Press 1 key to display only one method in the method selection list Press 0 key to keep t...

Page 308: ...range should be within these range to achieve best results Wavelength can be determined with Mode 53 Spectrum Abs E displayed Entering a User Concentration Press MODE 6 4 keys Confirm with key Enter Procedure The display shows Enter a method number in the range from 850 to 859 e g 8 5 0 Confirm with key Note if the entered number has already been used to save a concentration the display shows the ...

Page 309: ...ws Enter the concentration of the first standard e g 0 05 One step back with ESC Press F1 key to reset numerical input Confirm with key The display shows Prepare the first standard and press Test key The display shows the input value and the measured extinc tion value Confirm with key Enter the concentration of the second standard e g 0 1 One step back with ESC Press F1 key to reset numerical inpu...

Page 310: ... display shows the input value and the measured extinc tion value Confirm with key Note Perform as described above to measure further stan dards The minimum of measured standards is 2 The maximum of measured standards is 14 S1 to S14 If all required standards or the maximum value of 14 stan dards are measured press Store key The display shows The instrument goes back to the mode menu automati call...

Page 311: ...plus and minus sign Enter the exponent of the coefficient A e g 3 Confirm with key Successively the instrument queries the data for the other coefficients B C D E and F Note If zero 0 is entered for the value of the coefficient the input of the exponent is omitted automatically Confirm every input with key Enter measurement ranges from 0 05 to 2 5 Abs Press or key to change between plus and minus ...

Page 312: ...play shows The instrument goes back to the mode menu automati cally Now the polynomial is stored in the instrument and can be recalled by entering its method number or selecting it from the displayed method list TIP Save all your polynomial data in a written form because in case of power outage e g changing the battery all polynomial data will be lost and must be entered again You might want to us...

Page 313: ... selection list Press MODE 6 6 keys Confirm with key The display shows Enter the number of the User Method you want to delete in the range from 800 to 824 or 850 to 859 e g 800 Confirm with key There is displayed the query Press 1 key to delete the selected User Method Press 0 key to keep the valid User Method The instrument goes back to mode menu automatically User m clear choose no _______ 800 8...

Page 314: ...lynomial and concentration methods can be printed out or transferred with HyperTerminal to a PC Press MODE 6 7 keys Confirm with key The display shows Press key to print out the data e g wavelength unit of all stored User Methods The display shows e g After data transfer the photometer goes back to mode menu automatically M800 M803 User m print Start Mode 6 7 ...

Page 315: ...on to set it to a predefined state ATTENTION All stored user methods polynomial concentration are deleted with initialisation Press MODE 6 9 keys Confirm with key The display shows Confirm with key There is displayed the query Press 1 key to start initialisation Press 0 key to to cancel without initialisation The instrument goes back to mode menu automatically User m init Start Initialising YES 1 ...

Page 316: ...he display shows Enter the temperature value T in the range between 3 and 53 C and confirm with key If F was selected enter the temperature value in the range between 37 und 128 F The display shows Enter the value for Calcium hardness CH in the range between 50 and 1000 mg l CaCO3 and confirm with key The display shows Enter the value for Total Alkalinity TA in the range between 5 and 800 mg l CaC...

Page 317: ...splay message with key and enter a value in the defined range Selection of temperature unit Entering the temperature value is possible in degree Celsius or degree Fahrenheit Therefore the following preselection is once required Press MODE 7 1 keys Confirm with key The display shows Press 1 key to select degree Celsius Press 2 key to select degree Fahrenheit The instrument goes back to mode menu au...

Page 318: ...ess MODE 8 1 keys Confirm with key The display shows Press arrow key to increase brightness of the display about one unit Press arrow key to decrease brightness of the display about one unit Press arrow key Zero to increase brightness of the display about ten units Press arrow key Test to decrease brightness of the display about ten units The display shows The brightness can be selected between 0 ...

Page 319: ...t the current software versi on about the current detected mains power supply about the number of performed tests and free memory capacity Press arrow key to display the number of performed tests and free memory capacity Finish with ESC key System Info Software V012 002 3 003 002 more cancel Esc System Info Number of Tests 139 free records left 999 cancel Esc Mode 9 1 ...

Page 320: ...ce of the personal computer or printer using a cable in line with the specified assignment see technical data The cable for connection to a personal computer is included in delivery contents RS232 interface SP600 There are two RS232 interfaces at the photometer Interface F is for factory use only Interface D the operator needs to transfer data to a PC or printer or to perform an Update ...

Page 321: ...wing standard adjustments Detailed information of changing the adjustment you will find in the printer manual Baud rate 9600 Parity None Data bits 8 Note The printer must be connected and switched on before printing Caution Adjust printing parameter in Mode 29 See chapter 2 4 3 Printing Parameter 2 5 2 Data transfer to a personal computer Transferring test results from the photometer to a personal...

Page 322: ... applications and additional languages via internet Please find detailed information at our homepage on the download area The magnet part of the delivery content is needed for updating the instrument Remark To prevent loss of stored test results store or print out them before performing an Update ...

Page 323: ...cal distributor immediately 3 2 Delivery content Standard content of SP600 1 Photometer 2 batteries AA LR6 1 Power adapter 100 240 V 50 60 Hz 1 Cable for connection to PC 1 Magnet for updating SP600 1 Instruction manual 1 Manufacturer s Test Certificate M 1 Guarantee declaration Reagent sets and cells are not part of the standard scope of delivery Please see the General Catalog for details of avai...

Page 324: ...2 Abs at 635 nm 0 954 Abs x 0 994 Abs at 440 nm 0 907 Abs x 0 947 Abs at 635 nm Drift 0 005 Abs h at 500 nm Stray light 0 5 at 340 nm und 400 nm Wavelength accuracy 2 nm Wavelength reproducibility 1 nm Spectral band pass 10 nm Operation Acid and solvent resistant touch sensitive keyboard with integral beeper as acoustic indicator Power supply external power supply unit Input 100 240V 50 60Hz Outpu...

Page 325: ...itre KI Potassium Iodide KS 4 3 Acid demand to pH 4 3 this method is similar to the Total Alkalinity but converted into the unit mmol l as the German DIN 38409 demand TDS Total Dissolved Solids LR Low Range MR Medium Range HR High Range C Reagents of Chemetrics L Liquid reagent P Powder reagent PP Powder Pack T Tablet TT Tube Test DEHA N N Diethylhydroxylamine DPD Diethyl p phenylendiamine DTNB El...

Page 326: ...ithium battery Delete data with Mode 34 Jus Overrange E4 The user calibration is out of the accepted range Please check the standard reaction time and other possible faults Repeat the user calibration Jus Underrange E4 Overrange E1 The concentration of the standard is too high too low so that during user calibration the limit of the range was exceeded Perform test with a standard of lower higher c...

Page 327: ...are the same For this reason the combined Chlorine is most likely zero Example 2 The reading for free Chlorine is under the detection limit The instrument is not able to calculate the combined Chlorine In this case the combined Chlorine is most likely the same as the total Chlorine Example 3 The reading for total Chlorine is exceeding the range The instrument is not able to calculate the combined ...

Page 328: ...ne there is no selection between differentiated free or total Profi Mode is switched on Switch Profi Mode off with Mode 50 The pre programmed countdown is not displayed Countdown is not activated and or the Profi Mode is activated Switch the countdown on with Mode 13 and or switch the Profi Mode off with Mode 50 It seems that a method is not available Method is not activated in the user method lis...

Page 329: ... lamp Nevertheless from time to time it may be necessary to change this pre assembled component Please follow the instructions CAUTION Disconnect the instrument from the AC power and remember if instrument has been used recently the lamp can still be hot Note Don t touch the new halogen lamp L with the fingers since this would possibly interfere with a shelf life of the component 1 Disconnect the ...

Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...

Page 331: ...1 Switch the instrument off 2 Open the battery cover B 3 Remove the old batteries 4 Place the new batteries into the holder with the correct polarity 5 Put the cover back into position and lock it into place Caution Dispose batteries according to the local regulations ...

Page 332: ...people animals and goods security during the use of electrical materials which should be employed within certain voltage limits 73 23 CEE This conformity is presumed according to the following standard EN 61326 1997 A1 1998 A2 2001 A3 2003 When electrostatic discharge occurs close to the display or the metal parts in the cell chamber the display or the internal communication may be disturbed In th...

Page 333: ...331 SP600 Spectrophotometer 01 2008 ...

Page 334: ...Orbeco Hellige Inc 185 Marine Street Farmingdale NY 11735 Phone 631 293 4110 Fax 631 293 8258 service orbeco com www orbeco com ...

Reviews: