270
SP600 Spectrophotometer 01/2008
2.3 Operation mode
If the photometer is connected with power supply unit to the mains it is ready for use.
Before each start, make sure that the
sample chamber is empty
and the
photometer lid
is closed
, as the photometer always performs a selftest when it is switched on.
Switch the photometer on by pressing the ON/OFF key.
The display shows:
and then:
2.3.1 Selftest
The display shows:
get sure that the cell chamber is empty
and the photometer lid closed
Start selftest by pressing [ ] key.
• now the photometer is performing the selftest for
approx. 2
1
⁄
2
minutes. During this time the following
is checked:
• function of the tungsten halogen lamp
• function of the step motor
• accuracy of wavelength with internal Didymium filter
and if necessary adjustment (in this case the selftest can
be taken up to 5 minutes)
• function of data storage
When selftest is passed the list for method selection is dis-
played.
cell chamber empty?
Start: [ ]
selftest…
Initialisation…
Logo
PC Spectro II
or
Logo
SpectroDirect
ON
OFF
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...