
263
SP600 Spectrophotometer 01/2008
1.2.4 Sample dilution techniques
Proceed as follows for accurate dilutions:
Pipette the water sample (see table) into a 100 ml volumetric flask and fill up to 100 ml mark
with deionized water. Swirl to mix the contents.
Water sample
[ml]
Multiplication
factor
1
100
2
50
5
20
10
10
25
4
50
2
Pipette the required volume of the diluted sample into the vial and proceed as described in
the test methods.
Caution:
1. Dilution decreases accuracy.
2. Do not dilute water samples for measurement of pH-values. This will lead to incorrect
test results. If there is displayed “Overrange” use another instrument (e.g. pH-meter).
1.2.5 Correcting for volume additions
If a larger volume of acid or base is used to pre-adjust the pH-value, a volume correction of
the displayed result is necessary.
Example:
For adjusting the pH-value of a 100 ml water sample 5 ml of acid had to be added.
The corresponding displayed result is 10 mg/l.
Total volume
= 100 ml + 5 ml = 105 ml
Correction factor
= 105 ml / 100 ml = 1.05
Corrected result
= 10 mg/l x 1.05 = 10.5 mg/l
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...