203
SP600 Spectrophotometer 01/2008
1.1 Methods
12.
Fill a second clean vial with 10 ml of water sample.
13. Add
one GLYCINE tablet
straight from the foil and
crush the tablet using a clean stirring rod.
14. Close the vial tightly with the cap and swirl several
times until the tablet is dissolved.
15.
Transfer the contents of the second vial into the
prepared vial.
16. Close the vial tightly with the cap and swirl several
times until the tablets are dissolved.
17. Place the vial in the sample chamber making sure that
the marks are aligned.
18. Press
TEST
key.
Wait for a
reaction period of 2 minutes
.
After the reaction period is finished the measurement
starts automatically.
The result is shown in the display in:
mg/l Ozone
mg/l total Chlorine
Notes:
see page 197
Countdown
2:00
T1 accepted
prepare T2
press TEST
*,** mg/l O
3
*,** mg/l total Cl
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...