132
SP600 Spectrophotometer 01/2008
1.1 Methods
Hydrazine
with Powder Reagent
0.05 – 0.5 mg/l N
2
H
4
/ 50 – 500 μg/l N
2
H
4
1. Fill a clean vial (24 mm Ø) with
10 ml of the water sample
(Note 1, 2), close tightly with the cap.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
1 g HYDRAZINE test powder
(Note 3) to the
water sample.
6. Close the vial tightly with the cap and swirl several
times to mix the contents.
7. Press
[ ]
key.
Wait for a
reaction period of 10 minutes.
After reaction period is finished proceed as follows:
8. Any slight turbidity that occurs when the reagent is added
must be removed by filtration (Note 4).
9. Place the vial in the sample chamber making sure that
the marks are aligned.
10. Press
TEST
key.
The result is shown in the display as Hydrazine.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
10:00
start:
2
0
5
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...