295
SP600 Spectrophotometer 01/2008
Delete user-calibration
This chapter only applies for methods which can be user-calibrated.
Select the required method.
Instead of zeroing the instrument press [MODE] [4] [6] keys
and confirm with [ ] key.
The display shows:
Press [1] key to delete user-calibration.
Press [0] key to keep the valid user-calibration.
The instrument goes back to Zero-query automatically.
<user calibration>
100
Chlorine T
0.02-6 mg/l Cl2
clear user
calibration?
YES: 1, NO: 0
100
Chlorine T
0.02-6 mg/l Cl2
prepare ZERO
press ZERO
1
0
Mode
4
6