SP600 Spectrophotometer 01/2008
Spectrum (Scan)
A wavelength scan can be performed in the range of 330 to 900 nm.
The minimum range between start- and stop-wavelength is 10 nm.
Press [MODE] [5] [3] keys.
Confirm with
[ ]
key.
The display shows:
Enter start-wavelength:
e.g.: [4] [0] [0] and confirm with
[ ]
key.
Enter end-wavelength:
e.g.: [6] [2] [1] and confirm with
[ ]
key.
The display shows:
Place a filled cell (the blank) into the sample chamber (Note 1).
Press
ZERO
to start null balance (the baseline).
The display shows:
Place the cell with the test solution into the sample cham-
ber.
Press
TEST
key.
< Spectrum >
Start: _ _ _ nm
< Spectrum >
End: _ _ _ nm
< Spectrum >
400 – 621 nm
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
4 0 0
Mode
5
3
Test
Zero
6 2 1
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...