307
SP600 Spectrophotometer 01/2008
Confirm with
[ ]
key.
Press the appropriate numerical key to select the required
resolution.
Note:
Please enter the required resolution according to the instru-
ment presetting:
range
max. resolutions
0.000 ...9.999
0.001
10.00 ...99.99
0.01
100.0... 999.9
0.1
1000 ...9999
1
Measurement procedure with standards of
known concentration:
The display shows:
Prepare Zero and press [Zero] key.
Note:
Use deionized water or reagent blank value.
The display shows:
Enter the concentration of the first standard;
e.g.: 0.05
• One step back with [ESC].
• Press [F1] key to reset numerical input.
Confirm with
[ ]
key.
The display shows:
Prepare the first standard and press [Test] key.
The display shows the input value and the measured extinc-
tion value. Confirm with
[ ]
key.
Enter the concentration of the second standard;
e.g.: 0.1
• One step back with [ESC].
• Press [F1] key to reset numerical input.
Confirm with
[ ]
key.
< User concentr.>
prepare Zero
press ZERO
choose resolution
1: 1
2: 0.1
3: 0.01
4: 0.001
< User concentr.>
S1: 0.05 mg/l
prepare
press TEST
< User concentr.>
Zero accepted
S1: +________
| ESC | F1
S1 accepted
S2: +________
| ESC | F1
S1: 0.05 mg/l
E: 0.012
Test
Zero
0 .
0 5
0
.
1
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...