SP600 Spectrophotometer 01/2008
1.1 Methods
Ammonium HR
with Tube Test
1 – 50 mg/l N
1. Open one white capped reaction vial and add
0.1 ml
deionized water
(this is the blank).
2. Open another white capped reaction vial and add
0.1 ml of water sample
(this is the sample).
3. Add
one Vario AMMONIA Salicylate F5 powder
pack
straight from the foil into each vial.
4. Add
one Vario AMMONIA Cyanurate F5 powder
pack
straight from the foil into each vial.
5. Close the vials tightly and swirl several times to dissolve
the powder.
6. Press
[ ]
key.
Wait for a
reaction period of 20 minutes
.
After reaction period is finished proceed as follows:
7. Place the vial (the blank) in the sample chamber making
sure that the marks are aligned.
8. Press
ZERO
key.
9. Remove the vial from the sample chamber.
10. Place the vial (the sample) in the sample chamber
making sure that the marks are aligned.
11. Press
TEST
key.
The result is shown in the display in mg/l Ammonium.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
20:00
start:
1
2
3
4
Ø 16 mm
6
6
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...