107
SP600 Spectrophotometer 01/2008
1.1 Methods
Notes:
1. For determination of total Copper digestion is required.
2. Extremely acid water samples (pH 2 or less) must be adjusted between pH 4 and
pH 6 before the reagent is added (with 8 mol/l Potassium hydroxide solution KOH).
3. Accuracy is not affected by undissolved powder.
4. Interferences:
Cyanide, CN
-
Cyanide prevents full colour development.
Add 0.2 ml Formaldehyde to 10 ml of water sample and wait for a
reaction time of 4 minutes (Cyanide is masked). After this perform
test as described. Multiply the result by 1.02 to correct the sample
dilution by Formaldehyde.
Silver, Ag
+
If a turbidity remains and turns black, silver interferences is likely.
Add 10 drops of saturated Potassium chloride solution to 75 ml of
water sample. Filtrate through a fine filter. Use 10 ml of the filtered
water sample to perform test.
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...