159
SP600 Spectrophotometer 01/2008
1.1 Methods
Lead
with MERCK Spectroquant
®
Cell Test,
No. 1.14833.0001
0.1 – 5 mg/l Pb
Procedure B
1. Place the supplied blank in the sample chamber making
sure that the marks are aligned.
2. Press
ZERO
key.
3. Remove the vial from the sample chamber.
Caution! Reagent tubes contain Potassium Cyanide!
Adhere to the specified dosage sequence! (Note 4)
4. Fill one reaction tube with the same size drops by hol-
ding the bottle vertically and squeeze slowly:
5 drops reagent Pb-1K
5. Close the vial tightly with the cap and invert several
times to mix the contents.
6. Add
5 ml of water sample
.
7. Close the vial tightly with the cap and invert several
times to mix the contents.
8. Place the vial in the sample chamber making sure that
the marks are aligned.
9. Press
TEST
key.
10. Remove the vial from the sample chamber and open
carefully.
11. Add
1 level microspoon of reagent Pb-2K
.
12. Close the vial tightly with the cap and swirl until the
reagent is solved completely.
13. Place the vial in the sample chamber making sure that
the marks are aligned.
14. Press
TEST
key.
The result is shown in the display in mg/l Lead.
Ø 16 mm
prepare Zero
press ZERO
Zero accepted
prepare T1
press TEST
T1 accepted
prepare T2
press TEST
2
3
5
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...