128
SP600 Spectrophotometer 01/2008
1.1 Methods
Hardness, total HR
with Tablet
20 – 500 mg/l CaCO
3
1. Fill a clean vial (24 mm Ø) with
1 ml of the water sample
and
9 ml of deionized water
, close the vial with the
cap tightly.
2. Place the vial in the sample chamber making sure that
the marks are aligned.
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
one HARDCHECK P tablet
straight from the foil
to the water sample and crush the tablet using a clean
stirring rod.
6. Close the vial tightly with the cap and swirl gently
several times to dissolve the tablet.
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press
TEST
key.
Wait for a
reaction period of 5 minutes
.
After the reaction period is finished the measurement starts
automatically.
The result is shown in the display as total Hardness.
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Countdown
5:00
2
0
1
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...