292
SP600 Spectrophotometer 01/2008
3. Press
ZERO
key.
4. Remove the vial from the sample chamber.
5. Add
exactly 2 ml SPADNS reagent solution
to the
water sample.
Caution: Vial is filled up to the top!
6. Close the vial tightly with the cap and swirl gently
several times to mix the contents.
7. Place the vial in the sample chamber making sure that
the marks are aligned.
8. Press
TEST
key.
9. Remove the vial from the sample chamber. Empty the vial,
rinse vial and cap several times and then fill the vial with
exact
10 ml Fluoride standard
(Concentration 1 mg/l F).
10. Add
exactly 2 ml SPADNS reagent solution
to the
Fluoride standard.
Caution: Vial is filled up to the top!
11. Place the vial in the sample chamber making sure that the
marks are aligned.
12. Press
TEST
key.
The display shows:
Confirm with
[ ]
key.
Back to method selection with
ESC
key.
Select method Fluoride with keys [1][7][0] und
[ ]
.
Note:
The same batch of SPADNS reagent solution must be used
for adjustment and test. The adjustment process needs to
be performed for each new batch of SPANDS reagent so-
lution (see Standard methods 20th, 1998, APHA, AWWA,
WEF 4500 F D., S. 4-82).
As the test result is highly dependent on exact sample and
reagent volumes, the sample and reagent volumes should
always be metered by using a 10 ml resp. 2 ml volumetric
pipette (class A).
If there appears an error message please repeat adjust-
ment.
Zero accepted
T1: 0 mg/l F
press TEST
T1 accepted
T2: 1 mg/l F
press TEST
Calibration
accepted
Error, absorbance
T2>T1
1
7
0
Esc
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...