SP600 Spectrophotometer 01/2008
1.1 Methods
After reaction period is finished proceed as follows:
12. Place the vial
(the blank)
in the sample chamber
making sure that the marks are aligned.
13. Press
ZERO
key.
14. Remove the vial from the sample chamber.
15. Place the vial
(the sample)
in the sample chamber
making sure that the marks are aligned.
16. Press
TEST
key.
The result is shown in the display in mg/l Aluminium.
Notes:
1. Before using clean the vials and the measuring beaker with Hydrochloric acid
(approx. 20%). Rinse then thoroughly with deionized water.
2. To get accurate results the sample temperature must be between 20°C and 25°C.
3. A low test result may be given in the presence of Fluorides and Polyphosphates.
The effect of this is generally insignificant unless the water has fluoride added
artificially. In this case, the following table should be used:
Fluoride
[mg/l F]
Displayed value: Aluminium [mg/l Al]
0.05
0.10
0.15
0.20
0.25
0.30
0.2
0.05
0.11
0.16
0.21
0.27
0.32
0.4
0.06
0.11
0.17
0.23
0.28
0.34
0.6
0.06
0.12
0.18
0.24
0.30
0.37
0.8
0.06
0.13
0.20
0.26
0.32
0.40
1.0
0.07
0.13
0.21
0.28
0.36
0.45
1.5
0.09
0.20
0.29
0.37
0.48
---
Example: If the result of Aluminium determination is 0.15 mg/l Al and the Fluoride
concentration is known to be 0.4 mg/l F, the true concentration of Aluminium is
0.17 mg/l Al.
4. Al
Al
2
O
3
prepare Zero
press ZERO
Zero accepted
prepare Test
press TEST
Summary of Contents for SP600
Page 1: ...Instruction Manual SP600 Spectrophotometer ...
Page 2: ......
Page 9: ...7 SP600 Spectrophotometer 01 2008 Part 1 Methods ...
Page 37: ...35 SP600 Spectrophotometer 01 2008 ...
Page 83: ...81 SP600 Spectrophotometer 01 2008 ...
Page 163: ...161 SP600 Spectrophotometer 01 2008 1 1 Methods Note 1 Mn MnO4 KMnO4 ...
Page 203: ...201 SP600 Spectrophotometer 01 2008 ...
Page 207: ...205 SP600 Spectrophotometer 01 2008 ...
Page 249: ...247 SP600 Spectrophotometer 01 2008 1 1 Methods Notes 1 SO3 Na2 SO3 ...
Page 251: ...249 SP600 Spectrophotometer 01 2008 ...
Page 266: ...264 SP600 Spectrophotometer 01 2008 ...
Page 267: ...265 SP600 Spectrophotometer 01 2008 Part 2 Operating manual ...
Page 330: ...328 SP600 Spectrophotometer 01 2008 3 6 3 3 blank because of technical requirements ...